UPO 3000E&UPO2000E Series Oscilloscope Programming Manual
:CHANnel<n>:UNITs
Command format:
:CHANnel<n>:UNITs {VOLTs|AMPeres|WATTs|UNKNown}
:CHANnel<n>:UNITs?
Functional description:
Set channel unit as VOLTs, AMPeres, WATTs and UNKNown.
Return format:
Query return VOLTs, AMPeres, WATTs or UNKNown.
For example:
:CHAN1:UNIT VOLT set channel 1 unit as voltage
:CHAN1:UNIT? query return VOLTs
:CHANnel<n>:VERNier
Command format:
:CHANnel<n>:VERNier { {1|ON} | {0|OFF} }
:CHANnel<n>:VERNier?
Functional description:
Set position accommodation mode, ON presents Fine tuningto further
subdivides between coarse
tuning settings to improve vertical resolution; OFF presents Coarse tuning based on 1-2-5
system to set vertical sensitivity.
Return format:
Query return 1 or 0it presents ON or OFF.
For example:
:CHAN1:VERN ON turn on channel 1 fine tuning function
:CHAN1:VERN? query return 1
:CHANnel<n>:BIAS
Command format:
:CHANnel<n>:
BIAS { {1|ON} | {0|OFF} }
:CHANnel<n>:
BIAS?
Functional description:
The switch setting of channel bias voltage.
Return format:
Query return 1 or 0it presents ON or OFF.
For example:
:CHAN1:BIAS ON turn on channel 1 bias voltage function
:CHAN1: BIAS? query return 1