UNI-T
UPO 3000E&UPO2000E Series Oscilloscope Programming Manual
:MEASure:VPP?
Command format:
:MEASure:VPP? [<source>]
Functional description:
Measuring peak-to-peak value of the specified channel waveform, <source> take value as
CHANnel1, CHANnel2, CHANnel3 and CHANnel4. Ignore means the current channel.
Return format:
Query return 3.120e000unit is V.
:MEASure:VMAX?
Command format:
:MEASure:VMAX? [<source>]
Functional description:
Measuring the maximum value of the specified channel waveform, <source> take value as
CHANnel1, CHANnel2, CHANnel3 and CHANnel4. Ignore means the current channel.
Return format:
Query return2.120e000unit is V.
:MEASure:VMIN?
Command format:
:MEASure:VMIN? [<source>]
Functional description:
Measuring the minimum value of the specified channel waveform, <source> take value as
CHANnel1, CHANnel2, CHANnel3 and CHANnel4. Ignore means the current channel.
Return format:
Query return -2.120e000unit is V.
:MEASure:VAMPlitude?
Command format:
:MEASure:VAMPlitude? [<source>]
Functional description:
Measuring the amplitude value of the specified channel waveform, <source> take value as
CHANnel1, CHANnel2, CHANnel3 and CHANnel4. Ignore means the current channel.
Return format:
Query return3.120e000unit is V.