UPO 3000E&UPO2000E Series Oscilloscope Programming Manual
:TRIGger:VIDeo:STANdard?
Functional description:
Set the video standard.
Return format:
Query return{
NTSC | PAL | SECam }.
For example:
:TRIGger:VIDeo:STANdard NTSC set the video standard as NTSC
:TRIGger:VIDeo:STANdard? query return NTSC
:TRIGger:VIDEO:LINE
Command format:
:TRIGger:VIDEO:LINE <value>
:TRIGger:VIDEO:LINE?
Functional description:
Set the specified line of video sync, <value> presents the specified line and the range, which is
related to the video standard.
Return format:
Query return the specified line.
For example:
:TRIG:VIDEO:LINE 50 set the specified line of video sync as 50
:TRIG:VIDEO:LINE? query return 50
:TRIGger:VIDEO:SRATe?
Command format:
:TRIGger:VIDEO:SRATe?
Functional description:
To obtain the slew rate of video.
Return format:
Query return the current slew rate, return format is conform to binary data IEEE 488.2 # format.