UNI-T
UPO 3000E&UPO2000E Series Oscilloscope Programming Manual
SCPI Command
IEEE488.2 Common Command
*IDN?
Command format:
*IDN
Functional description:
For query manufacture name, model, product serial number and software version.
Return format:
Manufacture name, model, product serial number, software version separated by dot
mark.
For example:
UNI-T Technologies, UPO2000CS, UPO1000, 00.00.01
*RST
Command format:
*RST
Functional description:
Restore factory settings and clear the entire error message, send and receive queue buffers.
SYSTem Command
This command is for oscilloscope basic operation, including operating control, lock
full qwerty, error queue and system data.
:RUN
Command format:
:RUN
Functional description:
Start to sampling, execute :STOP command to stop it.
:STOP
Command format:
:STOP
Functional description:
Stop sampling, execute :RUN command to restart it.