Measurement Operation(4)
The resistance ranges are:
Model UT 58A/UT58B:200, 2k, 20k, 2M,20M and
200M.
Model UT 58C:200, 2k, 20k, 2M and 20M.
1. Insert the red test lead into the Hz V(UT58C)
or
V(UT58A/UT58B) terminal and the black test
lead into the COM terminal.
2. Set the rotary switch to an appropriate measurement
position in range.
3. Connect the test leads across with the object being
measured.
The measured value shows on the display.
black red
(figure 6)
Note
l The test leads can add 0.1to 0.2 of error to the slow-
resistance measurement. To obtain accurate readings
in low-resistance, short-circuit the input terminals
beforehand and record the reading obtained (called
this reading as X). (X) is the additional resistance from
the test lead. Then use the equation:
measured resistance value (Y) – (X) = accurate readings
of resistance.
l If the input terminal short-circuit reading 0.5,check
the test leads for any looseness or other cause.
l For high resistance (>1M), it is normal taking several
seconds to obtain a stable reading; select short test
leads for stable and precise readings.
l When the resistance is higher than the maximum range
or in open circuit condition, the Meter displays “1”.
l When resistance measurement has been completed,
disconnect the connection between the testing leads
and the circuit under test.
Model UT58A/B/C: OPERATING MANUAL