How to fix over-scale display on Unipulse Accessories?
- MmyerssamanthaAug 13, 2025
If the Unipulse Accessories display shows an over-scale reading, it could be due to the input signal from the load cell exceeding the F701-C span adjustment range. Verify that the load cell output is within the appropriate span adjustment range. Alternatively, the input signal from the load cell may be lower than the F701-C span adjustment range; in this case, check if the load cell output falls short of the span adjustment range.