What to do if the URIT Medical Electronic Measuring Instruments shows a low battery?
- RrobertscristianAug 14, 2025
If the screen displays a low battery warning in the right lower corner, replace the batteries.
What to do if the URIT Medical Electronic Measuring Instruments shows a low battery?
If the screen displays a low battery warning in the right lower corner, replace the batteries.
What to do if the URIT Medical Electronic PFS-30A displays an overdue test strip error?
If the test strip is overdue, first, check and correct the analyzer's current date and time, ensuring it is accurate. Second, ensure you are using the test strip within its valid date range.
What to do if the URIT Medical Electronic Measuring Instruments requires more blood?
If the blood volume is insufficient, ensure the volume of blood sample applied complies with the recommended volume in the test strip's instructions for use.
What to do if the URIT Medical Electronic PFS-30A Measuring Instruments gives a mismatched code card error?
If a mismatched code card is inserted, use the correct, matched code card for your URIT Medical Electronic Measuring Instruments.
What to do if the URIT Medical Electronic PFS-30A Measuring Instruments shows that the code card is not properly inserted?
If the code card is not properly inserted, remove it and re-insert the code card correctly into the URIT Medical Electronic Measuring Instruments.
What to do if I apply blood before the sampling indicator flashes on the URIT Medical Electronic Measuring Instruments?
If you apply the blood sample before the Sampling Indicator is flashing on the display, do not apply the blood sample on to the test strip until the Sampling Indicator is flashing.
What to do if the test strip is moved during automatic detection process of URIT Medical Electronic PFS-30A Measuring Instruments?
If the test strip is moved during the automatic detection process, ensure you do not move the test strip during this process.
What to do if the test hole is blocked by foreign substances before turning on the URIT Medical Electronic Measuring Instruments?
If the test hole is blocked by foreign substances, such as a test strip or fingers, before turning on the analyzer, move the foreign substances away and then turn on the analyzer.
How to clean the mirror surface of URIT Medical Electronic Measuring Instruments?
To clean the mirror surface, turn off the analyzer. Wipe the mirror surface clean using a dry and soft cloth or paper. Then, turn the analyzer back on.
Step-by-step guide for installing and replacing the AAA batteries in the analyzer.
Describes the functions and operation modes (short press, press and hold) of the analyzer's buttons.
Procedure for inserting the code card to match test strips for accurate readings.
Detailed steps for performing blood lipid tests, including preparation, sampling, and detection.
Important safety notes and precautions to be followed during the blood lipid detection process.
The URIT PFS-30/PFS-30A Blood Lipid Analyzer is a medical device designed for the quantitative detection of various blood lipid parameters in human whole blood samples. This analyzer is intended for use by both lay persons and healthcare professionals, making it suitable for daily blood lipid detection in individuals with normal or abnormal blood lipid levels. However, it is not suitable for neonatal testing, and its results should not be used for the sole diagnosis of cardiovascular diseases. The device operates based on reflection photometry, utilizing a built-in microcomputer to control and measure blood lipid content.
The primary function of the PFS-30/PFS-30A Blood Lipid Analyzer is to measure total cholesterol (TC), high-density lipoprotein cholesterol (HDL), triglyceride (TG), and low-density lipoprotein cholesterol (LDL) in human whole blood (venous or capillary blood) samples. Additionally, it calculates ratios such as TC/HDL based on these measurements. The analyzer is designed to work exclusively with the corresponding PFS 30 Blood Lipid Test Strips produced by URIT.
The measurement process involves inserting a test strip into the analyzer, which then automatically obtains a blank value. Once prompted, the operator applies a blood sample to the test area of the strip. Chemicals generated by the interaction of the blood sample and the test strip produce an absorption spectrum near 630nm wavelength, altering the optical reflection coefficient of the thin film on the test hole. The analyzer introduces light into the test zone and detects the intensity of the reflected light. A lower detected light intensity indicates a higher concentration of the detected substance, and vice versa. The internal computer then automatically calculates the blood lipid concentration based on the relationship between reflected light intensity and substance concentration, displaying the result on the LCD screen.
The analyzer also includes a CHD (Coronary Heart Disease) risk evaluation function, which can be initiated after successful detection. It offers two evaluation methods: FRA (Framingham CHD Risk Evaluation) and PRO (Procam CHD Risk Evaluation). The FRA method is used for evaluating CHD risk in the next 10 years for males or females aged 30 to 79 years old, while the PRO method is for males aged 35 to 75 years old. To perform a CHD risk evaluation, users input patient history details such as gender, age, smoking history, systolic blood pressure, and medical history for hypertension treatment. The analyzer then calculates and displays the CHD risk percentage.
The URIT PFS-30/PFS-30A analyzer is designed for user-friendly operation with several key features:
Proper maintenance is essential for ensuring the accuracy and longevity of the URIT PFS-30/PFS-30A Blood Lipid Analyzer:
Users must read the operation manual thoroughly before using the analyzer to prevent damage to the device or inaccurate results. Incorrect use can lead to fire or burns. The device is for external monitoring only and not for diagnosis. Always use the correct code card and test strips provided or recommended by the manufacturer. Do not disassemble the analyzer. Keep it away from children and pets. Avoid high temperature, humid environments, and contact with dust, salt, or sulfur. Do not immerse the analyzer in liquid or place it where it might fall into liquid. Do not place anything on top of the device or use it where aerosol droplets are present. Do not touch surfaces that are too hot. Avoid touching the sample area of the test strip. Do not use damaged or stained test strips or expired test strips. Do not press or move the analyzer or test strip during testing. Avoid direct exposure to strong light. Do not drop anything into the device's openings other than the test strip. The device complies with EN 61326-2-6-2013 and IEC 61010-1-2010 standards. Avoid using it near strong electromagnetic radiation sources. In medical institutions, evaluate the electromagnetic environment before use. Static discharges from synthetic materials can cause erroneous results. Venous blood tests should be performed by medical personnel following relevant operating rules. During detection, avoid touching buttons or the machine body. Wear protective gloves when handling blood samples. Ensure sufficient blood volume is applied the first time; do not reapply.
| Brand | URIT Medical Electronic |
|---|---|
| Model | PFS-30A |
| Category | Measuring Instruments |
| Language | English |