A
Ashley HoffmanAug 14, 2025
What to do if the URIT Medical Electronic Measuring Instruments shows a low battery?
- RrobertscristianAug 14, 2025
If the screen displays a low battery warning in the right lower corner, replace the batteries.
What to do if the URIT Medical Electronic Measuring Instruments shows a low battery?
If the screen displays a low battery warning in the right lower corner, replace the batteries.
What to do if the URIT Medical Electronic PFS-30A displays an overdue test strip error?
If the test strip is overdue, first, check and correct the analyzer's current date and time, ensuring it is accurate. Second, ensure you are using the test strip within its valid date range.
What to do if the URIT Medical Electronic Measuring Instruments requires more blood?
If the blood volume is insufficient, ensure the volume of blood sample applied complies with the recommended volume in the test strip's instructions for use.
What to do if the URIT Medical Electronic PFS-30A Measuring Instruments gives a mismatched code card error?
If a mismatched code card is inserted, use the correct, matched code card for your URIT Medical Electronic Measuring Instruments.
What to do if the URIT Medical Electronic PFS-30A Measuring Instruments shows that the code card is not properly inserted?
If the code card is not properly inserted, remove it and re-insert the code card correctly into the URIT Medical Electronic Measuring Instruments.
What to do if I apply blood before the sampling indicator flashes on the URIT Medical Electronic Measuring Instruments?
If you apply the blood sample before the Sampling Indicator is flashing on the display, do not apply the blood sample on to the test strip until the Sampling Indicator is flashing.
What to do if the test strip is moved during automatic detection process of URIT Medical Electronic PFS-30A Measuring Instruments?
If the test strip is moved during the automatic detection process, ensure you do not move the test strip during this process.
What to do if the test hole is blocked by foreign substances before turning on the URIT Medical Electronic Measuring Instruments?
If the test hole is blocked by foreign substances, such as a test strip or fingers, before turning on the analyzer, move the foreign substances away and then turn on the analyzer.
How to clean the mirror surface of URIT Medical Electronic Measuring Instruments?
To clean the mirror surface, turn off the analyzer. Wipe the mirror surface clean using a dry and soft cloth or paper. Then, turn the analyzer back on.
Brand | URIT Medical Electronic |
---|---|
Model | PFS-30A |
Category | Measuring Instruments |
Language | English |
Step-by-step guide for installing and replacing the AAA batteries in the analyzer.
Describes the functions and operation modes (short press, press and hold) of the analyzer's buttons.
Procedure for inserting the code card to match test strips for accurate readings.
Detailed steps for performing blood lipid tests, including preparation, sampling, and detection.
Important safety notes and precautions to be followed during the blood lipid detection process.