Temperature dependence :to.04 %RH/oC
Typical long-term stability better than 1 % RH per year
Response time 15 s with membrane filter
(at 20 °C, 90 % response)
Settling time 1 s
Sensor HUMICAP@ H-sensor
(part no. 0062)
Pt 100 (1/3 DIN 43760B,
four-wire connection)
7...35VDC
$:4 mA
-40 ...+60 °C
IP 55 (NEMA 3S)
IP 65 (NEMA 4) *
180g
radiation shield
Temperature sensor
Supply voltage
Current consumption
Operating temperature
Protection class
Weight
Accessories
HMP 35A
Sensor protection
.standard
.optional
membrane filter
(part no. 2787)
plastic grid (part no. 6597)
sintered filter 37 ~m
(part no. 6685)
sintered filter 216 ~m
(part no. 6686)
ASS plastic
1600 mm
Housing material
Cable length
HMP 35D
IEC 801-3 standard, level 2
E.M.I. protection
Sensor protection
.standard
membrane filter,
chromium-plated (part no.