43 42
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic envi-
ronment – guidance
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5% Uτ (>95%
dip in Uτ) for
½ cycle
40% Uτ (60%
dip in Uτ) for 5
cycles
70% Uτ (30%
dip in Uτ) for
25 cycles
<5% Uτ
(>95% dip in
Uτ) for 5 s
<5% Uτ (>95%
dip in Uτ) for
½ cycle
40% Uτ (60%
dip in Uτ) for 5
cycles
70% Uτ (30%
dip in Uτ) for
25 cycles
<5% Uτ
(>95% dip in
Uτ) for 5 s
Mains power quality
should be that of a
typical residential,
commercial or hospital
environment. The
battery powered
MIGHTY MEDIC
Vaporizer is able to offer
continuing functionality
even if the energy
supply is interrupted.
Note: Uτ is the a.c. mains voltage prior to application of the test level.
10. THE MIGHTY MEDIC VAPORIZER
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic envi-
ronment – guidance
Guidance and Manufacturer‘s Declaration – Electromagnetic Immu-
nity
The MIGHTY MEDIC Vaporizer is intended for use in the electromagnetic
environment specified below. The customer or the user of the MIGHTY MED-
IC Vaporizer should assure that it is used in such an environment.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V
rms
150 kHz to 80
MHz
3 V/m
80 MHz to 2,5
GHz
3 V
rms
3 V/m
Portable and mobile RF
communications
equipment should be
used no closer to any
part of the MIGHTY
MEDIC Vaporizer,
including cables, than
the recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
separation distance:
d = 1,2 √P
d = 1,2 √P
für 80 MHz to 800 MHz
d = 2,3 √P
for 800 MHz to 2,5 GH
10. THE MIGHTY MEDIC VAPORIZER