31
en
7.4.3. Foot Pedal
The micromotor can be started in 2 alternative
ways:
Either the foot pedal can be pressed and the
micromotor operates as long as the foot pedal
remains pressed,
or the micromotor may also be operated by
holding down the CONFIRM key (5), Fig. 4
for 1.5 seconds. The micromotor can be stopped
again by pressing any key or the foot pedal. To
conserve battery power, when operated with the
CONFIRM key , the micromotor will stop
automatically after 5 minutes of non-use.
7.4.4. Audible Signals
When pressing the keys on the keypad, a series
of audible tones and signals demonstrate that
the device is in use. All audible signals are active
(unless changed manually).
All audible signals are activated upon delivery of
the device:
Warning signal when torque exceeds approxi-
mately 75 % of preset value in rotary mode.
Warning signal when torque exceeds a critical
value in reciprocating mode.
• Intermittent signal when in continuous counter-
clockwise rotation.
Warning signal before the device switches off
when the battery is empty.
Deactivating and activating the audible signals
To deactivate the audible signals press the I and
I keys simultaneously: a beep will sound and
the following message will be displayed:
To reactivate the audible signals, press the I
and I keys simultaneously: a beep will sound
and the following message will be displayed:
7.4.5. File Library
After the R 1.0 software upgrade, the fi le library
will contain the following preset NiTi systems:
A) Reciprocating Systems
• RECIPROC
®
• WaveOne
TM
B) Rotary Systems
• Mtwo
®
(MTWO)
• FlexMaster
®
(FM)
Gates Glidden (GATE)
DR’S CHOICE (DR’S) individual program
• ProTaper
®
(PTU)
• K3
TM
The manufacturer reserves the right to update the
le library and the systems contained in it.
WARNINGS
Follow the file manufacturer’s in-
structions for use of endodontic files.
The file system shown on the display
must always match the file in use. This is
of the utmost importance in order to avoid
misusing reciprocating files and continuously
rotating files.
Torque and speed values are subject to
change by the file manufacturers without
notice. Therefore, the preset values in the
library must be checked prior to use. Torque
values shown on the display are accurate
and reliable only with VDW Endo 6:1 contra-
angles properly maintained and lubricated.
9.6.11_VDW.SILVER_RECIPROC Upgrade Stick_inhalt.indd 319.6.11_VDW.SILVER_RECIPROC Upgrade Stick_inhalt.indd 31 09.06.2011 17:18:0109.06.2011 17:18:01