Endo Easy Effi cient
®
33
en
7.5.3. ASR: Automatic Stop Reverse
NOTE
ASR function only works in rotary mode.
In reciprocating mode, ASR is disabled
(LED is off). Upon delivery the ASR function
is active (green LED). Select different
functions by pressing the ASR button.
In rotary mode, the ASR function
can be changed by pressing the ASR
button: and the LED color will change
according to the selected function:
Green: In rotary mode, when the preset torque is
reached, the micromotor will turn automatically in
reverse direction (counter-clockwise) until the file
no longer encounters resistance, at which point
it will automatically revert to forward rotation
(clockwise).
Red: In rotar y mode, the micromotor will turn auto-
matically in reverse direction (counter-clockwise)
with no torque control.
OFF: In rotary mode, when the preset torque is
reached, the micromotor will stop automatically.
By pressing the foot pedal again, the motor will
start in reverse direction (counter-clockwise)
until the file no longer encounters resistance, at
which point it will automatically revert to forward
rotation (clockwise).
Acoustic warning signals are given when:
• Torque exceeds approx. 75 % of preset value.
• In reverse rotation of the micromotor.
ASR function can be modified for each file, but the
setting will return to the default setting when the
file used will change.
7.6. Selecting a File System
The following chapters might also be helpful when
selecting a file system: description of keypad
(chapter 7.4.1.) and display (chapter 7.4.2.).
To choose a file system different from the one
currently displayed, the file system field (“sys”)
in the lower row must be active (indicated by an
arrow). If “sys” is not active, move the arrow to
the ”sys” field by pressing the I◄ and ►I keys
until it is active.
By pressing the + and - keys, you can scroll
through all systems until the display selects the
desired system. It is not necessary to confirm the
selection. The file system shown in the display is
the selected system. The display then reads, for
example:
7.6.1. Reciprocation
RECIPROC
®
and WaveOne
TM
instruments are
designed specifically for use in reciprocation,
whereby the instrument is driven first in a cutting
direction and then reverses to release the
instrument. The angles of reciprocation of the
motor are precise and specific to the design of
the instrument.
If one of the abovementioned reciprocating files
has been chosen, the display reading the first row
will show RECIPROCATION.
Select the specific reciprocating system
RECIPROC or WAVEONE by selecting the “system”
in the second row of the display via the + and - keys.
Contrary to the rotary mode, individual files not
have to be selected in the reciprocating mode, as
all motor settings are true for the entire system,
indicated by the word “ALL”.
9.6.11_VDW.SILVER_RECIPROC Upgrade Stick_inhalt.indd 339.6.11_VDW.SILVER_RECIPROC Upgrade Stick_inhalt.indd 33 09.06.2011 17:18:0209.06.2011 17:18:02