Endo Easy Effi cient
®
39
en
Problem Possible cause Solution
CALIBRATION
ERROR 1
An improperly connected
micromotor might disturb the
calibration process.
• Check that the micromotor is
correctly connected.
CALIBRATION
ERROR 2
The calibration process could
be disrupted by an excessively
resistant contra-angle.
• Check the contra-angle for any
malfunction.
• Lubricate the contra-angle carefully.
ROTARY FILE BLOCKS
IN CANAL
Wrong file setting.
Too much pressure on the
instrument.
• Switch to ASR mode “Reverse”
(LED red), start the motor and pull out
the file carefully.
RECIPROCATING FILE
BLOCKS IN CANAL
Too much pressure on the
instrument.
File not frequently cleaned.
• Try to remove the file with a pair of
pliers by pulling out and rotating the
file gently clockwise.
9.6.11_VDW.SILVER_RECIPROC Upgrade Stick_inhalt.indd 399.6.11_VDW.SILVER_RECIPROC Upgrade Stick_inhalt.indd 39 09.06.2011 17:18:0209.06.2011 17:18:02