? E115
l Several relays not set to the same failure mode are assigned
to the pump control
à All relays which are assigned to the pump control must be set
to the same failure mode
? E116
l Several relays are assignef to the pump control which are not
configured to the same mode
à All relays which are assigned to the pump control must be set
to the same mode
? E117
l A monitored pump signals failure
à Check the faulty pump. For acknowledgement, carry out the
reset "Failure relay 1 … 4" or switch the instrument OFF and
ON again
Depending on the failure reason and measures taken, the steps
described in chapter "Set up" must be carried out again, if necessary.
9.3 Instrument repair
If a repair is necessary, please proceed as follows:
You can download a return form (23 KB) from our Internet homepage
www.vega.com under: "Downloads - Forms and certificates - Repair
form".
By doing this you help us carry out the repair qu ickly and without
having to call back for needed information.
l Print and fill out one form per instrument
l Clean the instrument and pack it damage-proof
l Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
l Please ask the agency serving you for the address of your return
shipment. Y ou can find the respective agency on our website
www.vega.com under: "Company - VEGA worldwide"
Reaction after fault rec-
tification
54 VEGAMET 391 • 4 … 20 mA/HART signal conditioning instrument
9 Maintenance and fault rectification
36032-EN-091116