DC Current (Autorranging for uA and mA)
Range Resolution Accuracy
400.0uA 0.1uA +1.0% of rdg + 3 dgts
4000uA 1uA
40.00mA 10uA
400.0mA 100uA
+1.5% of rdg + 3 dgts
10A 10mA +2.5% of rdg + 5 dgts
Overload Protection: 0.5A / 250V and 10A / 250V fuse.
Maximum Input: 400mA DC or 400mA AC rms on uA / mA ranges, 10A DC or AC rms on 10A range.
AC Current (Autoranging for uA and mA)
Range Resolution Accuracy
400.0uA 0.1uA +1.5% of rdg + 5 dgts
4000uA 1uA
40.00mA 10uA
400.0mA 100uA
+1.8% of rdg + 5 dgts
10A 10mA +3.0% of rdg + 7 dgts
Overload Protection: 0.5A / 250V and 10A / 250V Fuse.
Frequency Range: 50 to 400 Hz
Maximum Input: 400mA DC or 400mA AC rms on uA / mA ranges, 10A DC or AC rms on 10A range.
Resistance (Autoranging)
Range Resolution Accuracy
400.0Ω 0.1Ω +1.2% of rdg + 4 dgts
4.000kΩ 1Ω +1.0% of rdg + 2 dgts
40.00kΩ 10Ω
400.0kΩ 100Ω
4.000MΩ 1kΩ
+1.2% of rdg + 2 dgts
40.00MΩ 10kΩ +2.0% of rdg + 3 dgts
Input Protection: 250VDC or 250VAC rms.
Capacitance (Autoranging)
Range Resolution Accuracy
4.000nF 1pF +5.0% of rdg+50 dgts
40.00nF 10pF +5.0% of rdg + 7 dgts
400.0nF 0.1nF
4.000uF 1nF
40.00uF 10nF
+3.0% of rdg + 5 dgts
200.0uF 0.1uF +5.0% of rdg + 5 dgts
Input Protection: 250VDC or 250VAC rms.