DVM9915
V. 03 29/11/2017 9 ©Velleman nv
ranging mode ......................................................... auto/manual
data hold ............................................................................ yes
backlight ............................................................................. yes
auto power-off ..................................................................... yes
dimensions ................................................... 188 x 102 x 38 mm
weight (with battery) ..................................................... ± 325 g
storage environment
temperature ................................................. -20 °C to 60 °C
humidity ............................................................ < 90 % RH
test lead probe (incl.) ............. CAT IV 600 V, 15 A; L = 90 cm (incl.)
IP rating ........................................................................... IP20
9.1 DC VOLTAGE
Do not measure circuits that may contain voltages
> 600 V
range
resolution
accuracy
400 mV
0.1 mV
± (0.5 % rdg + 2 digits)
4 V
1 mV
40 V
10 mV
400 V
100 mV
600 V
1 V
± (0.8 % rdg + 2 digits)
Overload protection: 600 V DC or AC rms
Impedance: 10 MΩ, > 100 MΩ on 400 mV range
9.2 AC VOLTAGE
Do not measure circuits that may contain voltages
> 600 V
range
resolution
accuracy
4 V
1 mV
± (0.8 % rdg + 3 digits)
40 V
10 mV
400 V
100 mV
600 V
1 V
± (1.2 % rdg + 3 digits)
Average sensing, calibrated to rms of sine wave
Frequency range: 40-500 Hz
Overload protection: 600 V DC or AC rms
Impedance: 10 MΩ