EasyManua.ls Logo

Velp Scientifica F201A0150 PW - User Manual

Velp Scientifica F201A0150 PW
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
1
Instruction Manual
Manuale di istruzioni
Manuel d’instructions
Manual de instrucciones
Bedienungsanleitung
PW - LH - LS - ES Overhead stirrers
F201A0150, F201A0156, F201A0151, F201A0152
General Information / Informazioni Generali / Informations Générales / Información General /
Allgemeine Hinweise
Before using the unit, please read the following instruction manual carefully.
Prima dell’utilizzo dello strumento si raccomanda di leggere attentamente il seguente manuale operativo.
Avant d’utiliser l’instrument, il est recommandé de lire attentivement le présent manuel d’instructions.
Antes de utilizar el instrumento, le recomendamos que lea con atención el siguiente manual de funcionamiento.
Bitte lesen Sie vor Inbetriebnahme des Geräts diese Bedienungsanleitung sorgfältig durch
Do not dispose of this equipment as urban waste, in accordance with EEC directive 2002/96/CE.
Non smaltire l’apparecchiatura come rifiuto urbano, secondo quanto previsto dalla Direttiva 2002/96/CE.
Ne pas recycler l’appareil comme déchet solide urbain, conformément à la Directive 2002/96/CE.
No tirar el aparato en los desechos urbanos, como exige la Directiva 2002/96/CE.
Dieses Gerät unterliegt der Richtlinie 2002/96/EG und darf nicht mit dem normalen Hausmüll entsorgt werden.
This unit must be used for laboratory applications only.
The manufacturer declines all responsibility for any use of the unit that does not comply with these instructions.
Questo strumento deve essere utilizzato solo per applicazioni di laboratorio.
La società produttrice declina ogni responsabilità sull’impiego non conforme alle istruzioni degli strumenti.
Cet instrument ne peut être utilisé que pour des applications de laboratoire.
Le fabriquant décline toute responsabilité en cas d’utilisation non conforme aux instructions concernant ces instruments.
Este dispositivo sólo debe utilizarse para aplicaciones de laboratorio.
El fabricante declina toda responsabilidad por el uso no conforme a las instrucciones de los dispositivos.
Dieses Gerät darf nur für Laboranwendungen verwendet werden.
Der Hersteller lehnt jede Haftung für unsachgemäße Verwendung oder Nichtbeachtung dieser Bedienungsanleitung ab.
This unit has been designed and manufactured in compliance with the following standards:
Lo strumento è stato progettato e costruito in accordo con le seguenti norme:
L’instrument a été conçu et fabriqué conformément aux normes suivantes:
El dispositivo se ha sido diseñado y fabricado de acuerdo con las siguientes normas:
Das Gerät wurde in Übereinstimmung mit folgenden Normen entwickelt und gebaut:
Safety requirements for electrical equipment for measurement, control and for laboratory use
Prescrizioni di sicurezza per apparecchi elettrici di misura, controllo e per l’utilizzo in laboratorio
Règles de sécurité pour appareils électriques de mesurage, de régulation et de laboratoire
Prescripciones de seguridad para equipos eléctricos de medición, control y su uso en laboratorio
Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte
IEC/EN 61010-1
IEC/EN 61010-2-051
Electrical equipment for laboratory use
UL 61010-1
General requirement - Canadian electrical code
CAN/CSA-C22.2 No.61010-1
Question and Answer IconNeed help?

Do you have a question about the Velp Scientifica F201A0150 PW and is the answer not in the manual?

Overview

This document describes the VELP Scientifica PW, LH, LS, and ES series of overhead stirrers, designed for laboratory applications requiring solubilization, mixing, suspension, emulsification, and homogenization of samples. These stirrers feature electronic speed control and advanced safety features, catering to diverse laboratory needs in terms of both viscosity and volume.

Function Description

The overhead stirrers are designed to provide controlled stirring for various laboratory processes. They utilize a keyless chuck for secure and easy mounting of stirring shafts up to 10mm in diameter. The stirring speed is electronically controlled, ensuring precise and consistent operation. The shaft bearings and belt drive contribute to silent running during operation.

Important Technical Specifications

The series includes four models: PW, LH, LS, and ES, each with distinct specifications:

F201A0150 PW:

  • Admitted power supply: 110-230V 50/60Hz (+/-10%)
  • Max. input/output power: 190 W / 435 W
  • Dimensions (WxHxD): 80x230x196 mm (3.1x9.0x7.7 in)
  • Weight: 2.9 Kg (6.4 lbs)
  • Speed range at nominal load: 20 - 1200 rpm
  • Speed indicator: Analogue
  • Max torque stirrer shaft: 120 Ncm
  • Max stirring quantity H2O: 70 litres
  • Clamping chuck clamping range: From 1 to 10mm
  • Overload and overtemperature protection: Automatic cut-off
  • Environmental ambient temperature: 5 - 40°C (41 - 104°F)
  • Storage temperature range: -10... + 60°C (14 - 140°F)
  • Max humidity: 80%
  • Noise Level: <80dBa
  • Operating mode: Continuous
  • Rapid Fuse diam. 5x20mm: 2xF1.6A L 250V

F201A0156 LH:

  • Admitted power supply: 110-230V 50/60Hz (+/-10%)
  • Max. input/output power: 190 W / 280 W
  • Dimensions (WxHxD): 80x230x196 mm (3.1x9.0x7.7 in)
  • Weight: 2.9 Kg (6.4 lbs)
  • Speed range at nominal load: 50 - 2000 rpm
  • Speed indicator: Analogue
  • Max torque stirrer shaft: 80 Ncm
  • Max stirring quantity H2O: 40 litres
  • Clamping chuck clamping range: From 1 to 10mm
  • Overload and overtemperature protection: Automatic cut-off
  • Environmental ambient temperature: 5 - 40°C (41 - 104°F)
  • Storage temperature range: -10... + 60°C (14 - 140°F)
  • Max humidity: 80%
  • Noise Level: <80dBa
  • Operating mode: Continuous
  • Rapid Fuse diam. 5x20mm: 2xF1.6A L 250V

F201A0151 LS:

  • Admitted power supply: 110-230V 50/60Hz (+/-10%)
  • Max. input/output power: 120 W / 180 W
  • Dimensions (WxHxD): 80x215x196 mm (3.1x8.5x7.7 in)
  • Weight: 2.3 Kg (5 lbs)
  • Speed range at nominal load: 50 - 2000 rpm
  • Speed indicator: Analogue
  • Max torque stirrer shaft: 40 Ncm
  • Max stirring quantity H2O: 25 litres
  • Clamping chuck clamping range: From 1 to 10mm
  • Overload and overtemperature protection: Automatic cut-off
  • Environmental ambient temperature: 5 - 40°C (41 - 104°F)
  • Storage temperature range: -10... + 60°C (14 - 140°F)
  • Max humidity: 80%
  • Noise Level: <80dBa
  • Operating mode: Continuous
  • Rapid Fuse diam. 5x20mm: 2xF1.6A L 250V

F201A0152 ES:

  • Admitted power supply: 110-230V 50/60Hz (+/-10%)
  • Max. input/output power: 30 W / 45 W
  • Dimensions (WxHxD): 80x160x200 mm (3.1x6.3x7.9 in)
  • Weight: 1.3 Kg (2.86 lbs)
  • Speed range at nominal load: 50 - 1300 rpm
  • Speed indicator: Analogue
  • Max torque stirrer shaft: 15 Ncm
  • Max stirring quantity H2O: 15 litres
  • Clamping chuck clamping range: From 1 to 10mm
  • Overload and overtemperature protection: Automatic cut-off
  • Environmental ambient temperature: 5 - 40°C (41 - 104°F)
  • Storage temperature range: -10... + 60°C (14 - 140°F)
  • Max humidity: 80%
  • Noise Level: <80dBa
  • Operating mode: Continuous
  • Rapid Fuse diam. 5x20mm: 2XF800mA L 250V

All models are designed for continuous operation and have a pollution degree of 2 and an overvoltage category of II, with a maximum installation altitude of 2000m (6561 ft) asl. The housing is made of die-cast polymer, the structure of aluminum, and the control panel of PET.

Usage Features

  1. Assembly and Installation:

    • The instrument should be placed on a flat laboratory bench, at least 30 cm from a wall.
    • It must be securely fastened to a support rod (A00001300) using a double clamp (A00001301).
    • The receptacle should be secured with a ribbon clamp (A00001302) to prevent damage.
    • The stirring shaft is assembled using an Allen head screw.
    • Before connecting to the power supply, ensure the voltage values on the rating plate match the power supply. The socket and cut-off device must conform to safety norms.
  2. Operating Controls:

    • The unit is turned on using the main ON/OFF switch.
    • Stirring speed is selected by turning the potentiometer knob, with speeds indicated on an analogue scale around the knob.
    • The selected speed setting is saved even after the instrument is turned off or in case of a power failure, and stirring resumes at the previously set speed upon restart.
  3. Overload Protection:

    • The stirrers feature electronic current and voltage limitation to protect the motor and electronic components from electrical surges.
    • In abnormal conditions, the unit automatically shuts down.
    • Over Load (Current surges): The stirrer can compensate for temporary current surges. If persistent, the motor speed gradually reduces to 0, and the unit shuts down.
    • Over Curr (Rpm error): This error occurs if the chuck is blocked or if the revs exceed the maximum permitted value.
    • Over Temp (Temperature error): This error occurs during prolonged heavy-duty work at ambient temperatures exceeding the maximum permitted value. The instrument must cool down before resetting.
    • Over Volt (Voltage surge): The software limits the motor's power supply to its nominal value (60V). High speed settings or heavy loads can cause higher voltage, which is then limited to 60V, reducing revs. If the motor's power supply exceeds 70V, the unit shuts down with an "OVERLOAD" message.
    • To reset the unit after an overload, remove the load from the chuck and cycle the power using the main switch.
  4. Safety Precautions:

    • The plug is the disconnect means; ensure the instrument is placed where it can be quickly disconnected.
    • Do not use the instrument in explosive environments or with dangerous/flammable substances.
    • Do not immerse the unit in water.
    • Always place the stirring shaft in the receptacle before turning the unit on to avoid dangerous free-air rotation.
    • Users are responsible for decontaminating the instrument if hazardous substances fall on or in it. Use safety substances for cleaning that do not react with internal parts.
    • If the solution being stirred may release toxic, dangerous, or poisonous gases, adequate safety measures (e.g., hood, personal protective equipment) must be taken.
    • The working speed should be set to avoid wobbling and/or splashes.

Maintenance Features

  1. Routine Maintenance:

    • No routine or extraordinary maintenance is required beyond periodic cleaning.
    • Any repairs must be carried out by VELP Scientifica or authorized local distributors, in accordance with product guarantee laws.
    • The instrument should be transported in its original packaging, following any specific instructions (e.g., palletized).
  2. Cleaning:

    • Disconnect the unit from the power supply before cleaning.
    • Use a cloth dampened with a mild, non-flammable detergent.
    • Users are responsible for proper decontamination of the unit if hazardous substances are present on its surface or interior. Contact the manufacturer or distributor if unsure about cleaning product compatibility.

Accessories / Spare Parts

A range of accessories and spare parts are available, including:

  • A00001300: Support rod and base
  • A00001301: Double clamp
  • A00001302: Ribbon clamp
  • A00001304: Stirring shaft with floating blade
  • A00001305: Stirring shaft with folding blade
  • A00001306: Stirring shaft with fixed blade
  • A00001307: Stirring shaft with propeller
  • A00001308: Stirring shaft with paddle, 6 holes
  • A00001309: Stirring shaft with turbine
  • A00001310: Stirring shaft with turbo propeller
  • A00001311: Stirring shaft with anchor
  • 10005266: Knob 26D blue
  • 10000132: Gasket for overhead stirrer

The instruments comply with IEC/EN 61010-1, IEC/EN 61010-2-051, UL 61010-1, and CAN/CSA-C22.2 No.61010-1 safety standards, and directives 2006/42/EC, 2014/35/EU, 2014/30/EU, 2011/65/UE (RoHS), and 2012/19/UE (RAEE).

Velp Scientifica F201A0150 PW Specifications

General IconGeneral
BrandVelp Scientifica
ModelF201A0150 PW
CategoryLaboratory Equipment
LanguageEnglish