Ventana BenchMark GX Quick Reference Guide
Basic Slide Run
I. Prepare the Instrument
A. Empty the waste container
1. Pull out the bottle and un-screw the top
2. Empty the bottle according to laboratory
procedures.
3. Replace the bottle, screw on the top,
and push the bottle into place.
B. Relleachbulkbottle,asneeded:
4. To remove a bulk bottle, press the latch
release and pull the bottle out
C. Toinstallabulkbottle:
5. Rellthebottlewiththesameproduct
6. Place the bottle in the slot and push it
forward until you feel or hear it click.
Thebottlettingshouldbeseated
securelyagainsttheinstrumenttting.
Positions of Bulk Products on the Instrument
Bulk Solution
Container ID Number Bulk Solution ContainerGraduated Carboy
05244676001
Bulk Solution Bottle
2L
EZ PREP
EZ PREP
05279755001
Liquid Coverslip
05264839001
SSC
05353947001
Reaction Buffer
05353955001
CC1
05279801001
Concentrates
CC2
05279798001
Option
Bottle
20L
Instrument Bulks
H2O DIS
2L
2L
3L 3L 3L 3L6L 3L3L
2L 2L 2L 2L 2L
Liquid Coverslip
2L
SSC
20L
10L
H2O DIS
2L
Reaction Buffer
20L
H2O DIS
2L
CC1
2L 2L
1L
CC2
1.
1.
4.