- 4 -
This scale had been calibrated in factory before shipment, however, from
time to time, there is a remarkable change in temperature or humidity
changed, there might be inaccuracies, in this case, there needs to
re-calibrate the scale again.
When the scales is powered off, press the “ON/OFF”key until to show: PE
xxx, release it and immediately, press both “ON/OFF” and “MODE” keys at
the same time until to show: E – 100, Press “MODE” key again will show: E
– 50, E – 25, E – 10, E – 5. Place a Standard calibration weight: 100kg or
50kg or 25kg or 10kg Or 5kg on the middle of the platform according to the
LED shown digit,(For example: E – 10, place a standard 10 kgs calibration
weight), until The LED shows: “Good”, calibration is over.
Attention: more weight, more accuracy after calibration.
Display show: H” to mean: the scale has exceeded the weight limited
(over loaded)
Display show: “-E” to mean: the batteries power are too low, it needs to
replace the Batteries or be Operated by adaptor.