12
11. Technical Specications
Type: Forehead Thermometer V977C-AUS
Measuring Range:
Forehead measuring range: 34°C to 42.2°C.
Resolution:
0.1°C
Accuracy:
Laboratory: ±0.2°C, 36.0 ~ 39.0°C
(Ambient Temperature: 16°C to 40°C)
Display:
Liquid Crystal Display, 4 digits plus special icons
Acoustic:
a. The unit is turned ON and ready for the measurement:
1 short «beep» sound.
b. Complete the measurement: 1 long beep sound.
c. System error or malfunction: 3 short «beep» sounds.
d. Temperature higher than 37.4°C (yellow or red screen
display): 10 beeps
Memory:
8 readings recall in the Memory Mode.
Backlight
a. The display will be lighted when the unit is turned ON and
will stay lit until the power button is pressed to take a
measurement.
b. The display will be lighted again for 15 seconds when the
measurement has been completed.
Operating temperature:
16°C to 40°C
Storage/
Transport Temperature:
-20°C to +50°C
Storage/
Transport Humidity:
15-95% Non-Condensing
Automatic Switch-off:
Approx. 30 seconds after last measurement has been taken
Battery:
(2) AAA Batteries - at least 1000 measurements
Weight:
3.51oz. (with battery), 2.72oz. (w/o battery)
Standards:
Complies with ASTM E-1965-98 requirements
P/N: 31IMV977292 DATE: FEB14
Type BF Equipment
12. Warranty
This product has been produced with the greatest care for normal, household use. Following
directions carefully will ensure dependable operation. If for any reason (other than misuse or
normal battery operation) you are dissatised with your Vicks Forehead Thermometer, we will
repair or replace it, at our option, at no charge during the 3 year warranty period. Modication
to the product by the consumer is not authorised and voids the warranty.
If the thermometer does not function properly, rst check the battery. Replace if necessary.
Batteries and packaging are excluded from the warranty.
If you experience unsatisfactory operation, please contact Felton Grimwade & Bosisto’s. If
repair is required under the warranty period, the product should be returned with proof of
purchase to Felton Grimwade & Bosisto’s service centre.
The purchaser assumes responsibility for the proper care and use of the product in accordance
with the printed operating manual. The purchaser or user must make his or her judgment as
to when to use the product and the length of use. Keep the operation manual available as it is
your guide to safe, efcient operation.
Our goods come with guarantees that cannot be excluded under the Australian Consumer Law.
You are entitled to a replacement or refund for a major failure and for compensation for any
other reasonably foreseeable loss or damage. You are also entitled to have the goods repaired
or replaced if the goods fail to be of acceptable quality and the failure does not amount to a
major failure.
Manufacturing Quality Healthcare Products for
Over 75 Years
This product is manufactured by Kaz USA,
Inc. under license from The Procter &
Gamble Company. Vicks and other associated
trademarks are owned by The Procter & Gamble
Company. Fever InSight is a registered trademark
of Kaz USA, Inc.
© 2013 Kaz USA, Inc.
Made and printed in China
www.kaz.com/vicks
Distributed by: Felton Grimwade & Bosisto’s
Pty. Ltd.
61-81 Clarinda Road (Po Box 74),
Oakleigh South, Victoria 3167, Australia
Free call: 1800 655 841
E-mail: mail@fgb.com.au
Website: www.fgb.com.au

Other manuals for Vicks V977C-AUS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vicks V977C-AUS and is the answer not in the manual?

Vicks V977C-AUS Specifications

General IconGeneral
BrandVicks
ModelV977C-AUS
CategoryThermometer
LanguageEnglish

Related product manuals