14
DESCRIPTION: the projection/display shows “Lo.5”.
PROBLEM: the room temperature is too low (below 10,0°C/50.0°F).
SOLUTION: move to another, warmer room and, if you are taking
the body temperature, wait for the stabilization of the device and of
the subject
Messages about the detected temperature
DESCRIPTION: the projection/display shows
«Hi.2» and the value alternately.
MEANING: ATTENTION! The temperature is
40,0°C/104.0°F or more.
DESCRIPTION: the projection/display reads “Hi.2”.
PROBLEM:
- reading with “face”
button: the temperature detected exceeds
the limit for operation with this button (>43,0°C / >109.4ºF).
- reading with “home”
button: the temperature detected exceeds
the limit for unit operation (>80,0°C / >176.0°F).
SOLUTION:
- reading with “face”
button: please make certain that you’ve
pressed the correct button, with regard to the measured object, and
that the warnings have been met.
- Reading with “home”
button: the temperature cannot be taken.
DESCRIPTION: the projection/display shows “Lo.3”.
PROBLEM:
- reading with “face”
button: the forehead temperature appears to
be too low (<34,0°C / <93.2°F).
- reading with “home”
button: the surface temperature read is
below the operating limit (<1.0°C / <33.8°F).
SOLUTION:
- reading with “face”
button: make certain the sensor is not soiled
or damaged and that the subject has not come from a cold room.
- reading with “home”
button: the temperature reading cannot
be taken.
DESCRIPTION: the display shows «o».
PROBLEM: the button that you’ve pressed is disabled when the de
-
vice is set in NURS.
SOLUTION: if you want to enable it. see par. #7.