9. MEANING of DISPLAY MESSAGES
11 - English -
DESCRIPTION: during normal operation, the symbol appears on the display.
PROBLEM: the batteries are running low but it is still possible to take several readings.
SOLUTION: obtain new batteries for changing them when the signal E.1 (below)
appears.
DESCRIPTION: “E.1” and the symbol appear on the display, or the unit does not
turn on at all.
PROBLEM: the batteries are dead.
SOLUTION: remove the batteries immediately and insert the new ones when
necessary (see par. #11).
DESCRIPTION: the projection/display reads "E.8".
PROBLEM: the thermometer was moved before the light started flashing, or the area
is subject to strong electromagnetic fields.
SOLUTION: wait until the lights flash before moving the thermometer; make certain
that you are not in the vicinity of calls with mobile or cordless telephones.
DESCRIPTION: the display reads C:AL and/or a countdown (in
minutes and seconds), a pictogram of a hand inviting to stop, and
the messages AQCS and MQCS?
PROBLEM: the thermometer has not stabilized.
SOLUTION: • wait until the AQCS countdown has run its course
without touching the thermometer, or
• perform an MQCS (par. #8).
DESCRIPTION: while in stand-by mode, the room temperature
displayed is accompanied by the symbol MQCS or AQCS.
MEANING: the thermometer has undergone manual or automatic
quick calibration (par. #8).
DESCRIPTION: if you’ve pressed the “FACE” button and
the projection/display shows "Hi.4" and the value alternately.
PROBLEM: the room temperature is between 40.1 and 45°C
(104.1 and 113°F).
MEANING: the temperature reading can be taken but accuracy
is not guaranteed.
DESCRIPTION: the projection/display reads "Hi.4”.
PROBLEM: the room temperature is too high (above 45°C/113°F).
SOLUTION: move to another, cooler site and, if you are taking the body temperature,
wait for the stabilization of the device and of the subject.
DESCRIPTION: the projection/display shows "Lo.5" and the
value alternately.
MEANING: - reading with “FACE” button: the room
temperature is between 10 and 15.9°C (50 and 60.6ºF).
- reading with “HOME” button (DOCT mode): the room
temperature is between 41 and 60.6ºF (5 and 15.9°C).
SOLUTION: the temperature reading can be taken but
accuracy is not guaranteed.