Accuracy
°C
from 1,0 to 19,9°C:
from 20,0 to 35,9°C:
from 36,0 to 39,0°C:
from 39,1 to 42,5°C:
from 42,6 to 80°C:
±
1,0°C
±
0,3°C
±
0,2°C
±
0,3°C
±
1,0°C
Accuracy
°F
from 33.8 to 67.9°F
from 68.0 to 96.7°F
from 96.8 to 102.2 °F
from 102.3 to 108.6°F
from 108.7 to 176 °F
±
1.8°F
±
0.5°F
±
0.4°F
±
0.5°F
±
1.8°F
13. TECHNICAL CHARACTERISTICS
Series VisioFocus
®
, model VisioFocus
®
PRO 06480: non-contact infrared thermometer
Resolution: 0.1
Body temperature readings Measurement range: 34,0/42,5°C (93.2-
108.5
°F)
Room temperature working range: 16/40°C (60.8/113°F )
(1)
from 34,0 to 35,9°C:
±0,3°C
from 36,0 to 39,0°C: ±0,2°C
(2)
from 39,1 to 42,5°C:
±0,3°C
Other readings
Measurement range: 1,0/80,0°C (33.8-176°F)
Room temperature working range: 10/40°C (40/104°F)
(3)
(1)
In rooms where the temperature is between 10 and 15,9°C (50 and 60.6°F), accuracy and the operating range
are not guaranteed and the message “Lo.5" and the temperature value are displayed alternately. In rooms where the
temperature is between 40,1 and 45°C (104.1 and 113°F), accuracy and the operating range are not guaranteed
and the message “Hi.4" and the temperature value are displayed alternately.
(2)
ASTM E1965-98-2009 laboratory accuracy requirements in the display range of 37 to 39°C (98 to 102°F) for IR
thermometers is
±
0,2 °C (
±
0.4 °F), whereas for mercuryin-glass and digital thermometers, the requirement per ASTM
Standards E 667-86 and E 1112-86 is
±
0,1 °C (
±
0.2°F).
3)
In rooms where the temperature is between 5 and 9,9 °C (41 and 50°F) accuracy and the operating range are not
guaranteed and the display shows "Lo.5" and the value alternately. In rooms where the temperature is between 40,1
and 45°C (104.1 and 113°F), accuracy and the operating range are not guaranteed and the message “Hi.4" and the
temperature value are displayed alternately.
Power supply: 4 AAA (LR03) alkaline batteries - 1.5 V (included)
Life of high quality batteries: up to 3 years or 30,000 readings (depending on use)
Dimensions: 144 x 43,5 x 21,5 mm (5.67 x 1.71 x 0.85 inches) - including cap
Weight: 98 gr. (3.46 oz.) - batteries included
Distance from the subject: calculated using an optical aiming system (approx. 6 cm/2.36 inches).
Large, clearly visible, backlit display in 5 colors:
1. light blue (button used: FACE. See par. #4.1 and #4.2);
2. green (button used: HOME. See par. #4.3);
3. violet (button used: MEM, memory function. See par. #5);
4. orange (button used: MEM, ambient temperature function. See par. #6);
5. blue (MQCS function. See par. #8).
Atmospheric pressure range of operating conditions: from 700 hPa to 1,060 hPa. Relative humidity
range of operating conditions: from 15% to 93%, non-condensing.
Keep in a clean, dry place, preferably at a temperature ranging between +16 and +40°C (60.8 and
104°F). Store at a temperature included between -10 and +60°C (14 and 140°F) and in any case
not lower than -18°C (0.4°F) or higher than +70°C (15F).
Expected life: 10 years.
VisioFocus PRO is a class IIa medical device (per Directive 93/42/EEC and subsequent amendments
and integrations). Its measuring system has been tested in hospitals, private clinics and medical
offices. It is compliant with the pertinent ASTM standard (E1965-98:2009) and with IEC 60601-1
and IEC 60601-1-2 standards. Internally powered equipment for continuous operation. The
VisioFocus PRO LEDs emit low light radiation in compliance with IEC 62471. Tecnimed srl, P.le
Cocchi, 12 - 21040 Vedano Olona (VA) - Italy undertakes full responsibility for this product's
compliance with the reference standards.
Accuracy
°C
Accuracy
°F
from 93.2 to 96.7°F:
±0.5°F
from 96.8 to 102.2°F: ±0.4°F
(2)
from 102.3 to 108.5: ±0.5°F