EasyManua.ls Logo

VistaLab MLA Series User Manual

VistaLab MLA Series
2 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
Operators Manual
MLA Digital Pipettes
Introduction
MLA Digital Pipettes are to deliver air displacement adjustable volume instruments. Three
models are available covering a volume range of 10 to 1000µL. Volume is adjusted by rotating
the plunger. The plunger clicks in place for any desired setting. The volume display is located
in the body of the pipette.
Pipette Tips
It is recommended that MLA Pipettes be used with MLA Pipette Tips. The use of tips from
other sources may degrade the pipette performance. For information on MLA Pipette Tips
refer to the Pipette Information Table.
Pipetting Procedure
a. Rotate the plunger to display the desired volume (clockwise to decrease and
counterclockwise to increase).
b. Using MLA Pipette Tips, press the pipette nozzle firmly into a fresh tip.
c. Fully depress the pipette plunger and then immerse the tip into the solution (approximately
1/8 inch-3mm deep).
d. Smoothly release the plunger and allow the solution to enter the pipette tip.
e. Remove the tip from the solution and touch the tip against the side of the vessel to remove
any solution that may have adhered to the outside of the tip.
f. Place the tip against the side of the receiving vessel as close to the bottom as possible or,
if the vessel contains liquid, as close to the liquid as possible. Smoothly depress the plunger.
g. While holding the plunger depressed, slowly withdraw the tip keeping it against the wall of
the container.
h. Release the plunger and remove the tip by applying a slight upward pressure to the bonnet.
(See Figure 1)
VistaLab Technologies, Inc.
27 Radio Circle Drive
Mt. Kisco, NY 10549 USA
(888) 652-6520
www.vistalab.com
326720 Rev. C
Volume Range Catalog No. Color Increment
10 to 50µL 1136 Green 0.1µL
50 to 200µL 1137 Red 0.2µL
200 to 1000µL 1138 Gray 1.0µL
Specifications Table
Volume Volume Change Volume Acceptable Accuracy
per 1/2 revolution
Setting Range (µL)
10 to 50µL 2.5µL Low 10µL 9.5 - 10.5 ±5%
Mid 30µL 29.1 - 30.9 ±3%
High 50µL 49.0 - 51.0 ±2%
50 to 200µL 5.0µL Low 50µL 49.0 - 51.0 ±2%
Mid 125µL 123.75 - 126.25 ±1%
High 200µL 198.0 - 202.0 ±1%
200 to 1000µL 25µL Low 200µL 198.0 - 202.0 ±1%
Mid 600µL 594.0 - 606.0 ±1%
High 1000µL 990.0 - 1010.0 ±1%
Pipette Information Table
MLA Pipette Tips
Volume Color Seal Kit Stacked Bulk Sterile
Rack
10 to 50µL Green 9091
9025
4025 2025
50 to 200µL Red 9092
4225 2027
200 to 1000µL Gray 9094 9026 4026 2026
4226 2028
Accessories
Nozzle Inserts 8066
Pipette Stand 1700
pipette
Question and Answer IconNeed help?

Do you have a question about the VistaLab MLA Series and is the answer not in the manual?

VistaLab MLA Series Specifications

General IconGeneral
BrandVistaLab
ModelMLA Series
CategoryLaboratory Equipment
LanguageEnglish

Summary

Introduction to MLA Digital Pipettes

MLA Pipette Tips

Recommendation to use MLA Pipette Tips for optimal pipette performance.

Pipetting Procedure

Pipetting Steps

Detailed steps for aspirating and dispensing liquids using MLA Digital Pipettes.

Pipette Handling Tips

Pre-wetting Tips

Pre-wetting tips with sample liquid to improve accuracy with certain fluids.

Plunger Control

Smooth and consistent plunger action is crucial for accurate pipetting.

Air Bubble Prevention

Tips to prevent air bubbles during liquid aspiration for accurate volumes.

Pipette Handling

Advice on maintaining verticality and consistent tip immersion depth.

Nozzle and Insert Maintenance

Instructions for cleaning and checking nozzle inserts and assemblies.

Pipette Calibration

Calibration Procedure

Detailed steps for testing and adjusting pipette volume accuracy.

Pipette Maintenance

Lubrication and Cleaning

Guidelines for routine lubrication and cleaning of pipette internal parts.

Seal Check and Replacement

Procedure for checking seals for wear and instructions for replacement.

Related product manuals