CorVitals, Distributor
Electromagnetic Emission Declaration
The VitalPatch device is intended for use in the electromagnetic environment specified below. The end user of the
device should assure that it is used in such an environment.
Measurements
ECG Dynamic Range
-10mV to +10mV
Heart Rate (stationary
and ambulatory)
30 – 200 Beats per Minute (<±5 or 10% Beats per Minute,
whichever is greater)
Respiration Rate
o
10-30 Breaths per Minute with a mean absolute error of less
than 3 Breaths per Minute, validated by clinical studies
o
4-42 Breaths per Minute with a mean absolute error of less
than 1.5 Breaths per Minute, validated by simulation studies
Skin Temperature
15
0
C – 50
0
C (± 0.3
0
C )
Fall Detection
Fall or No Fall (> 90% Sensitivity and >98% Specificity)
Step Count
< 5% Absolute Error Compared to Manual Count
Step count is reset to 0 after step count 65535 is reached.
Posture Detection
Lying down, Upright, Walking, Running, or Leaning (>70% Accuracy
Compared to Visual)
Communications
Bluetooth (BT4.1)
Max. 10 Meters (30 Feet Line of Sight)
Radio Modulation
FSK (Frequency Shift Keying)
Radio Frequency
2.4 – 2.5GHz
Transmit power
10dbm
Security
AES-CCM 128 Bit Encryption (Advanced Encryption Standard-CCM
mode)
Battery
Battery Type
Zinc Air
Battery Voltage
DC 1.4V
Battery Life
120 Hours
Operating
Conditions
Ambient Temperature
10 – 40
o
C
Humidity
10 – 95% RH
Altitude
<3000 m
Barometric Pressure
70 kPa to 102 kPa
Emission test
Compliance
Electromagnetic environment
RF emissions
CISPR 11
Group 1
The VitalPatch device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The VitalPatch device is suitable for use in all establishments,
including domestic establishments and those directly connected to
the public low-voltage power supply network that supplies buildings
used for domestic purposes.