22
The Choice of Power Supply
1.Battery powered mode: 6VDC 4*AA batteries
Installing and Replacing the Batteries
1. Open the battery door.
2. Insert the batteries according to the polarity indications.
3. Close the battery door.
Battery Life: Approx. 70 days
(Battery capacity: 600 mAH. If measured twice per day, each measurement
takes 35s, measuring result display takes 60s. The current for
measurement is 350 mA and that for records display is 35 mA, while the
current when shutdown is 25 uA.
In order to achieve the best performance and protect you monitor,
please use the authorized / specified battery.
CAUTION
Replace the batteries under following circumstances:
displays on the LCD.
The LCD display dims.
When powering on the monitor, the LCD doesn’t light up.
Remove batteries if the device is not likely to be used for some time.
Worn batteries are harmful to the environment. Do not dispose with daily garbage.
Remove the old batteries from the device following your local recycling guidelines
.
CAUTION