VOLTECH AT3600 USER MANUAL PROGRAM MODE
Three of the options require more than just pressing the third soft-key.
For ‘Beeper Volume’, pressing the soft-key CHANGE brings up an additional
display:
• Press the ↑ and ↓ soft-keys until the beeper makes the level of sound you require,
then press the EXIT soft-key to return to the Set-Up list.
For ‘Display Contrast’, the additional display is very similar, again with ↑ and ↓ soft-
keys, which this time adjust the display contrast.
For the ‘Clock’ option, there are five fields to set up:
• Initially the cursor will be in the first field - ‘yyyy’ - which is the year. Type the
number you require using the keyboard, and press either the ENTER key ( ↵ ), or
the → soft-key to move to the next field.
• For the second field - ‘mmm’ - which is the month, the soft-keys change to give
the names of each month. Press the appropriate soft-key to select which month.
• The remaining fields - ‘dd’, ‘hh’ and ‘mm’ (day, hour and minute) are changed in
the same way as the year, using the number pad.
To reset the set-up options to the factory defaults:
Press the decimal point (“.”) on the instrument keypad during boot-up at the general
information screen. This will reset all tester options to the factory defaults, including
screen contrast and beep volume, excluding the real-time clock.
The factory default settings are:
Option Default Setting
FAIL Beeper OFF
PASS Beeper OFF
Key Response CLICK
Beeper Volume 6
Display Contrast 8
Program Store INTERNAL
FRONT PANEL OPERATION VPN 98 - 024 PAGE 6.6.3.