Disinfection Sensors
Instruction Manual

Questions and Answers

  • B
    barbara69Sep 23, 2025
    What causes 'this device can work faster' message with no Local Area Connection 2 on Walchem Controller?
    • C
      christinecontrerasSep 23, 2025
      If you receive the message 'this device can work faster' but no Local Area Connection 2 appears on your Walchem Controller, it suggests that the driver in your PC has crashed. This can occur when the Controller Restart button is clicked with the USB cable connected. Reboot your PC. If a blue screen appears on restart attempt, a hard reboot may be necessary. Alternatively, the driver in your PC may be corrupt; in this case, uninstall and reinstall the driver.
  • M
    melissakellyAug 23, 2025
    What to do if there is a faulty controller sensor input in Walchem WDSW100?
    • K
      Kim FlowersAug 25, 2025
      If you suspect a faulty controller sensor input in your Walchem Accessories, start by going to the Sensor Input Page and running a self-test. If the test passes, the issue likely lies with the sensor or its wiring. If the self-test fails, disconnect the sensor from the circuit board and try the self-test again. If failure persists, consider replacing the circuit board.
  • C
    Christopher CareyAug 25, 2025
    What to do if there is no electrolyte fill solution in the membrane cap of Walchem Accessories?
    • M
      Melinda LamAug 26, 2025
      If there's no electrolyte fill solution in the membrane cap of your Walchem Accessories, refill the membrane cap with electrolyte. If the cap won't hold the solution, replace it.
  • V
    vbauerAug 27, 2025
    What to do if there are air bubbles on the membrane of Walchem Accessories?
    • S
      Steven FlynnAug 27, 2025
      If you notice air bubbles on the membrane of your Walchem Accessories, dislodge the bubbles. If necessary, increase the flow rate.
  • C
    cschmidtAug 27, 2025
    What to do if there is insufficient sample flow in Walchem WDSW100?
    • S
      Sharon TaylorAug 27, 2025
      If you have insufficient sample flow in your Walchem Accessories, increase the flow rate to between 30 and 100 liters per hour.
  • C
    Catherine ZimmermanSep 18, 2025
    Why does the 'New Hardware Found Wizard' appear when connecting my Walchem Controller?
    • C
      cordovablakeSep 18, 2025
      If you encounter the 'New Hardware Found Wizard' after connecting the cable to your Walchem Controller, it could be due to the following reasons: * The driver is not installed: Install the driver. * The driver is installed on another USB port: Move the cable to the other port.
  • E
    Edward BradleySep 15, 2025
    What to do if I can't reconnect my Walchem Controller after unplugging the cable?
    • T
      Tricia MurphySep 15, 2025
      If you are unable to reconnect your Walchem Controller after removing and re-plugging the cable, the driver in your PC may not have reset. To resolve this, unplug the cable and wait for 5 minutes, or reboot your PC.
  • S
    stevenharrisSep 18, 2025
    Why does my Walchem Controller show 'page cannot be displayed' after entering the IP address?
    • T
      Thomas FosterSep 18, 2025
      If you get the message 'this device can work faster' and Local Area Connection 2 appears, but entering 199.199.199.2 in the browser results in 'The page cannot be displayed' on your Walchem Controller, it means that Local Area Connection 2 is not set to use a fixed IP address. To fix this, access the ‘Local Area Connection 2’ properties and enter the fixed IP address of 199.199.199.1 as described in the Quick Start Guide.
  • M
    msandovalSep 6, 2025
    Why are my Walchem WDSW100 Accessories not sending emails even though ShoulderTap® is working?
    • J
      John WoodsSep 6, 2025
      If your Walchem Accessories are not sending emails but ShoulderTap® is working, the issue could stem from several factors: * The SMTP IP address might be incorrect, or not for an ASMTP server. Verify this information with your email service provider. * Ensure that 'Use ASMTP' is checked and that the ASMTP username and password are correct. * Confirm you have a valid email account set up and that the controller's email address is accurately entered. * Also, verify that the 'Send Email Via' setting is configured to 'Modem'. * Finally, double-check that the recipient email address is correct, and there isn't a general email server problem. Contact your email service provider to check the SMTP IP address.
  • R
    Robert WhiteSep 3, 2025
    Why does calibration fail due to slope calculated in One Point Process Calibration being outside of the range in Walchem WDSW100 Accessories?
    • R
      rothjamesSep 3, 2025
      A calibration failure will occur in Walchem Accessories if the slope (mV/ppm) calculated in the One Point Process Calibration is outside of the range of the nominal mV per 0.1 to 2.0 ppm.