Directions for Use 45
Nellcor
®
Sensor Accuracy Guide
Accuracy specifications are based on controlled hypoxia studies with healthy,
non-smoking adult volunteers over the specified saturation SpO
2
range. Pulse oximeter
SpO
2
readings were compared to SaO
2
values of drawn blood samples measured by
hemoximetry. All accuracies are expressed as + “X” digits. This variation equals plus one
standard deviation (+ 1 SD), which encompasses 68% of the population.
Nellcor Patents
Covered by one or more of the following US Patents: 4,802,486; 4,869,254; 4,928,692;
4,934,372; 5,078,136; 5,351,685; 5,485,847; 5,533,507; 5,577,500; 5,803,910; 5,853,364;
5,865,736; 6,083,172; 6,708,049; Re. 35,122 and foreign equivalents.
Pulse Rate Range 25 to 240 bpm
Pulse Rate Accuracy ±3 bpm
±3 bpm (low perfusion)
Table 15. OxiMax Sensor Models, Single Patient Use
Sensor Models SpO
2
Range
70% to 100%
MAX-AI ± 2
MAX-PI ± 2
MAX-II ± 2
MAX-RI
1
± 3.5
1
The accuracy specification has been determined between saturations of 80% to 100%.
Table 16. OxiCliq Sensor Models, Single Patient Use
Sensor Models SpO
2
Range
70% to 100%
OXICLIQ-PI ± 2.5
Table 17. Reusable Sensor Models
Sensor Models SpO
2
Range
70% to 100%
D-YS (Infant to Adult) ± 3
D-YS and D-YSE ± 3.5
D-YS and D-YSPD ± 3.5
DS-100A ± 3
OXI-A/N (Adult) Adult: ± 3
OXI-P/I (Pediatric/infant) ± 3