Directions for Use Chapter 2 Setup 19
Use the settings menu to select and set the following parameters:
To change a settings parameter:
1. Select the parameter as indicated above.
2. Change the value by pressing or .
3. Set the displayed new value either by doing nothing for 10 seconds or by pressing any
button other than or . If you press a function button (such as ), the monitor
returns to normal operation with that function ( ) activated.
Patient Type Neonate Term birth through 28 days, or up to 44 gestational weeks
Pediatric 29 days through 12 years
Adult 13 years and older
Target Pressure The initial cuff inflation pressure (set individually for each patient type)
Temp Modes Fahrenheit Predictive
Fahrenheit Monitored
Celsius Predictive
Celsius Monitored
Pulse Tone Volume From 0 (silent) to 5 (loudest)
ºF
ºF
ºF
ºF
M
M
ºC
ºC
ºC
ºC
M
M