23
Calibration Data Write Error occurred
/
24
Invalid Present Bar Operation due to a
Calibration Data Error /
Description/Probable Cause
Conf. Register Error / if non-recoverable re-calibrate
Config,
Register then replace Microprocessor U34.
Halt in Response to Remote Command
/
if non-recoverable
possible defective Audio/Tymp Communications Chip U36
or Microprocessor
U34.
Invalid
Keycode
detected in CMDTSK
/
if non-recoverable
possible defective Switch Panel, Switch Buffer U26, U27, or
U28, or Microprocessor U34.
Invalid
Keysource detected in CMDTSK
/
same as above.
Invalid
above.
Queue Power
Error detected in CMDTSK /same as
Bad Pointer to EEPROM Routine’/
Collision of the Normal Stack and Swap
Stack
/
Invalid function for a Banded Swapped Function
/
Printer RAM Error on power up
/
internal or micro
Tymp Hardware State is unknown
/
Non-Fatal Audiometer related error messages
Calibration Data Read Error
occured
/
HL Limit exceeded /
Invalid Extended Range Selection due to HL not in range
where the Extend Range is valid
/