16
key to turn on/off the buzzer, the buzzer icon on the LCD is turned
on/off accordingly. After setting, press SET one time to save the
setting, if no input/key operation in 5 seconds, the setting is autoly
saved and the device autoly enters body temperature mode.
Display and Alarms
Body temperature reading:
1)Body temperature reading range:32℃~37.3℃(89.6℉~99.1℉)
display: green, alarm: none;
a. 2)Body temperature reading range:37.4℃~38℃(99.2℉~100.4℉)
display: yellow, alarm: 3 continuous “Beep sound;
b. 3)Body temperature reading range:38.1℃~42℃(99.2℉~100.4℉)
display: red, alarm: 6 continuous “Beep sound;
1) 5)In body temperature mode, if the measured reading is lower than
32℃, it displays “Lo, if reading is higher than 42℃, it displays “HI”.
2) 4)If the measuring result shows reading higher than the preset F3
alarming value, the LCD backlight turns red and the device sounds 6
continuous “Beep sound.
3) 6) The end of the measurement data stay display data, the
screen is always bright for easy recording.
4) 7) when the battery voltage is lower than 2.8V, the low battery icon
is shown.
5) 8)The buzzer icon appears if the buzzer is set to be on, the icon
disappears if the buzzer is set to be off.
6) 9)Long press the MODE key for 3 seconds to enter the memory data
log mode, “M” is displayed on LCD.
7) 10)If no input/key operation over 15 seconds, the device autoly turns off.
7. Read Before Use
Before measuring forehead temperature, make sure there is no hair, sweat,
makeup, hat or anything else covering the targeted position and the
thermometer is pointed perpendicularly to the middle of forehead.
To ensure measuring accuracy, it is recommended that the thermometer
and the person to be measured should be in a room at normal room
temperature for at least 30 minutes prior to the measurement.
In order to obtain the highest accuracy in successive measurements,
please wait for 30 seconds between two measurements.
After sleeping, is recommended to wait for 5 minutes before measuring
the body temperature.
When the person being measured is from a place where the temperature
thereof is materially different from the temperature of the environment
which the measuring process takes place, such person should stay in the
measuring environment for at least 5 minutes before the measuring
process starts.
Cold compress on the forehead of a feverish patient or other cooling
measures alike is very likely to compromise the accuracy of the
measurement, avoid measuring under such circumstances.
The ambient temperature of the person being measured should remain
stable, do not measure body temperature in places with strong airflow
such as in front of a fan or air vents of air conditioners.
Do not measure body temperature directly after showering, swimming