Adjusting the temperature parameters for the temperature alarm (AL and AH)
The temperature alarm is pre-set to sound, as soon as a temperature of 1 ℃ or less is reached.
Setting the low temperature alarm (AL)
1. Push and hold the settings button for 3 seconds.
2. Push the “+” or “–” button until “AL” is shown on the temperature display.
3. Push the settings button.
4. Push “+” or “–” to set the alarm as required.
Setting the high temperature alarm (AH)
1. Push and hold the settings button for 3 seconds.
2. Push the “+” or “–” button until “AH” is shown on the temperature display.
3. Push the settings button.
4. Push “+” or “–” to set the alarm as required.
Adjusting the parameters for the air humidity alarm (AS)
The temperature alarm is pre-set to sound, as soon as a temperature of 45 % is reached. This is suffi-
cient and does not need to be altered.
Setting the low humidity alarm (AS)
1. Push and hold the settings button for 3 seconds.
2. Push the “+” or “–” button until “AS” is shown on the temperature display.
3. Push the settings button.
4. Push “+” or “–” to set the alarm as required.
By filling water into both water channels, the air humidity should rise by 60 %, depending on the local
humidity and season of year. It is recommended to fill the water channels up every 4–5 days. On day
18, when the egg inserts are removed, they should be overfilled, to raise the humidity to approx. 65 %.
Make sure that neither the eggs, nor the equipment are allowed to get wet.
Calibration of the temperature measurement (CS)
The correct measured value of the thermometer is set to 0 ℃. The displayed value of the thermometer
can be adjusted if you realise that the temperature value is wrong with a calibrated thermometer.
Calibration of the temperature sensor measurement
1. Push and hold the settings button for three seconds
2. Push the “+” or “–” buttons, until “AH” is displayed on the temperature display.
3. Push the settings button.
4. Push “+” or “–” to set the alarm as required.
Make sure that the adjustments are the deviation between the thermometer values and should be
adjusted with the “–” if the temperature of the incubator is too high and the normal value (display +
value) is too low.
Adjusting the upper and lower temperature limits (HS and LS)
HS (high setting) and LS (low setting) adjust the limits of the adjustment range of the desired tempera-
ture (incubation temperature regulation).
If HS is set to 38.2 ℃ and LS to 37.4 ℃, the desired temperature (adjustment of the incubation tem-
perature) can only be altered from 38.2 ℃ to 37.4 ℃, thus the minimum temperature is limited to
37.2 ℃, even if the “–” button is pushed down and held. The same applies to the high setting. This
disables an accidental change of temperature outside of the required range.