DOLPHINN Steam Sterilizer Operation Manual
Rev-A Page 11 of 17
4.2 Specification
Basic specification
Rated Voltage: AC 220V/ 50~60Hz/ 4A, or AC 110V/ 50~60Hz/ 8A if request.
Rated Power: 1600VA
Fuse: T10A AC220V /T25A AC110V
Operation Temperature: 5~40℃
The Board Affording Weight: 4000 N/m
2
Noise :< 50db
The Max. capacity of a plate:1000g
The Max. duration of using loading test: 90mins.
The Max. thermal radiation energy under the condition of 20℃~26℃:<2000J.
Sterilizer Chamber
Material: stainless steel 304 (for medical)
Work Pressure: -0.9 bar~2.5 bar
Max. Temperature: 145 ℃
Chamber Volume: 18L(Φ245×350mm)
Loading Size: 18L (537×417×410)
Max. Loading Weight: 18L (3.07kg/cm
2
)
Working Pressure/Temperature: 1.1~1.3bar/ 121~122℃; 2.1~2.3bar/ 134~135℃
Water Volume for One Cycle: 0.16L~0.18L
Sterilizer Steam Safety Valve
Safety Release Pressure: 2.45 bar
Max. Working Temperature: 160 ℃