5.4. Load the trays into the chamber with tray removal tool, and put the instruments which need
to be sterilized on the trays.
5.5. Close the door.
5.6. Turn on the power switch, then set the sterilization temperature, temperature select switch
ON: 134℃/ 0.22Mpa/ 6mins, temperature selection OFF:121℃/ 0.12Mpa/ 25mins. It will
start sterilization cycle automatically.
5.7. When the cycle is finished, the green lamp will be kept on. Then please turn on the air
exhausting switch to release the pressure inside the chamber to 0 bar, which takes about 6
to 8mins. Then you can open the door.