Oxi 330/SET
Oxi 340/SET
Sauerstoff-Taschenmeßgerät
Dissolved Oxygen Hand-Held Meter
Oxi 330
Oxi 340
Sauerstoffsensor
Dissolved Oxygen Probe
CellOx 325
Luftkalibriergefäß
Air Calibration Beaker
OxiCal
®
- SL
ba22211de

Table of Contents

Questions and Answers

  • J
    Jason SimpsonAug 9, 2025
    Why is the measured value too high on my wtw Oxi 330?
    • C
      Charles ReidAug 9, 2025
      If the measured value from your wtw Oxygen Equipment is too high, it could be due to a high share of solutes or air bubbles hitting the membrane with high velocity. Correct the solubility function by entering the salinity equivalent. Ensure the flow doesn't directly meet the membrane. Note that if the carbon dioxide pressure is too high (over 1 Bar), no measurement is possible.
  • N
    Nicholas QuinnAug 17, 2025
    What to do if my wtw Oxi 330 shows inadmissible calibration?
    • J
      jerrywatsonAug 17, 2025
      If your wtw Oxygen Equipment shows an inadmissible calibration, it might be due to several reasons: * The electrolyte solution may be used up. * The membrane could be contaminated. * The electrode system may be poisoned. In such cases, regenerate the probe or clean the membrane.
  • C
    Christopher SparksAug 25, 2025
    Why is the measured value too low on my wtw Oxi 330?
    • V
      Vanessa MasonAug 25, 2025
      If the measured value from your wtw Oxygen Equipment is too low, it might be due to insufficient flow. Ensure adequate flow is provided.
  • W
    Wendy MorganSep 1, 2025
    What to do if the indication range is exceeded on wtw Oxi 330?
    • W
      whitejonathanSep 1, 2025
      If your wtw Oxygen Equipment's indication range is exceeded, it could be due to a short circuit between the gold and lead electrodes. Check the probe and exchange it if necessary.

Summary

Safety Guidelines

Control Elements and Display

Keypad Operation and Functions

Explains the function of each button on the device keypad.

Putting the Device into Operation

Battery Operation and Replacement

Instructions for using and replacing batteries in non-rechargeable versions.

Mains and Rechargeable Battery Operation

Details AC power connection and operation with rechargeable batteries.

Device Calibration Procedures

Measuring Oxygen with Salinity Correction

Details how to apply salinity correction for accurate oxygen concentration readings.

Probe Supervision and Symbols

Drift Control AR Functionality

Data Logging and Memory Management

Analog Output Functionality

Serial Data Transmission

Device Configuration Settings

Reset Functionality

Troubleshooting Common Issues

Oxi System Messages and Solutions

Explains system error codes, their causes, and troubleshooting steps.

Other O2 Measurement Errors

Addresses specific problems that may occur during oxygen measurements.

Technical Data and Specifications

CellOx 325 Dissolved Oxygen Probe Manual

Probe Troubleshooting Guide

Provides solutions for common problems and error symptoms encountered with the probe.

wtw Oxi 330 Specifications

General IconGeneral
Brandwtw
ModelOxi 330
CategoryOxygen Equipment
LanguageEnglish