Table of Contents

Questions and Answers

  • M
    Marco YangAug 9, 2025
    Why is the measured value too high on my wtw OxiCal-SL Oxygen Equipment?
    • F
      Frances RodriguezAug 9, 2025
      If the measured value from your wtw Oxygen Equipment is too high, it could be due to a high share of solutes or air bubbles hitting the membrane with high velocity. Correct the solubility function by entering the salinity equivalent. Ensure the flow doesn't directly meet the membrane. Note that if the carbon dioxide pressure is too high (over 1 Bar), no measurement is possible.
  • J
    Jeremy BakerAug 17, 2025
    What to do if my wtw OxiCal-SL Oxygen Equipment shows inadmissible calibration?
    • A
      aguilarjoyAug 17, 2025
      If your wtw Oxygen Equipment shows an inadmissible calibration, it might be due to several reasons: * The electrolyte solution may be used up. * The membrane could be contaminated. * The electrode system may be poisoned. In such cases, regenerate the probe or clean the membrane.
  • W
    Wayne SaundersAug 25, 2025
    Why is the measured value too low on my wtw Oxygen Equipment?
    • J
      Jesse SkinnerAug 25, 2025
      If the measured value from your wtw Oxygen Equipment is too low, it might be due to insufficient flow. Ensure adequate flow is provided.
  • L
    Leslie FlemingAug 31, 2025
    What to do if the indication range is exceeded on wtw OxiCal-SL Oxygen Equipment?
    • B
      Brian WoodSep 1, 2025
      If your wtw Oxygen Equipment's indication range is exceeded, it could be due to a short circuit between the gold and lead electrodes. Check the probe and exchange it if necessary.

Summary

Safety Guidelines

General Safety Precautions

Details crucial safety precautions and warnings for operating the instrument safely, including electrical and environmental considerations.

Oxi 330;340 Instruction Manual

Instrument Control Elements

Describes the instrument's buttons, display icons, and status indicators for operation and navigation.

Device Setup and Initial Operation

Covers battery installation, mains connection, and initial device setup for first-time use.

Oxygen Sensor Calibration Procedures

Explains how to calibrate the oxygen sensor and set essential measurement parameters.

Measured Value Storage and Data Logging

Covers storing measured values, manual/automatic saving, memory full indication, and read-out operations.

Troubleshooting: System Messages and Errors

Provides solutions for common system messages and errors encountered during O2 measurement.

CellOx 325 Dissolved Oxygen Probe Manual

Galvanic Oxygen Sensor Regeneration Report

wtw OxiCal-SL Specifications

General IconGeneral
Brandwtw
ModelOxiCal-SL
CategoryOxygen Equipment
LanguageEnglish