Oxi 330
Oxi 340
Troubleshooting
44
D. O. system messages
System message Cause Action
/
Indication range
exceeded
Short circuit between gold
and lead electrode
Check probe,
if necessary exchange
Inadmissible
calibration
•
Electrolyte solution used up
•
Membrane contaminated
•
Electrode system poisoned
Regenerate probe
Clean membrane
Regenerate probe
Damaged
membrane
•
Membrane damaged
•
Membrane head not
tightened properly
Regenerate probe
Tighten membrane head
Other possible errors - D. O. measurement
System message Cause Action
Measured value
too low
Insufficient flow Provide flow
Measured value
too high
•
High share of solutes
•
Air bubbles bump on
membrane with high velocit
•
The carbon dioxide pressure
is too high > 1 Bar
Correct solubility function by
entering the salinity equivalent
Flow should not meet
membrane directly
No measurement possible