EasyManua.ls Logo

wtw OxiCal-SL - Troubleshooting; D. O. System Messages; Other Possible Errors - D.O. Measurement

Default Icon
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Oxi 330
Oxi 340
Troubleshooting
44
D. O. system messages
System message Cause Action
/
Indication range
exceeded
Short circuit between gold
and lead electrode
Check probe,
if necessary exchange

Inadmissible
calibration
Electrolyte solution used up
Membrane contaminated
Electrode system poisoned
Regenerate probe
Clean membrane
Regenerate probe

Damaged
membrane
Membrane damaged
Membrane head not
tightened properly
Regenerate probe
Tighten membrane head
Other possible errors - D. O. measurement
System message Cause Action
Measured value
too low
Insufficient flow Provide flow
Measured value
too high
High share of solutes
Air bubbles bump on
membrane with high velocit
y
The carbon dioxide pressure
is too high > 1 Bar
Correct solubility function by
entering the salinity equivalent
Flow should not meet
membrane directly
No measurement possible

Table of Contents

Related product manuals