Troubleshooting
Oxi 330
Oxi 340
45
General system messages
System message Cause Action
Meter does not
react on
keypressing
Undefined operation state
of the processor, e. g.
after inadmissible EMC
admission (outside the
specified ranges)
Reset of processor:
Press RCL key and
switch meter on.
LoBat
Batteries nearly discharged.
Charge rechargeable batteries/
Change non-rechargeable
batteries
*(
Timeout of serial interface. Check connected
instrument.
is flashing
Set calibration interval has
run out.
Recalibrate probe.