DISPLAYING MESSAGES AND TROUBLESHOOTING
As listed below are problems that may appear with your GLM-76 Blood Glucose Meter when
performing a test. These messages will help you to identify certain problems. If you have any
concerns or see any other error display, please contact us or the local distributor for help.
NOTE:
If there is any error message
displayed but not listed here,
or your meter shuts down for
no reason, you can reset the
meter or replace the battery.
DISPLAY
DESCRIPTION
ACTION TO TAKE
D
isplay check
I
f some parts of the display do not work Contact your local distributor for
h
elp
Blood glucose result may be
h
igher than 33.3 mmol/L or 60
0
m
g/dL.
If this is not confirmed by the way you feel, review proper testing
procedure and perform a quality check with control solution. Repeat
blood test, if “HI” still appears, call your doctor’s immediately.
Blood glucose result may be
lower than 1.1 /L or 20 mg/dL.
If this is not confirmed by the way you feel, review proper testing procedure
and perform a quality check with control solution. Repeat blood test. If “LO”
still appears, call your doctor’s immediately
T
emperature is above or belo
w
t
he operating range of test strips
The result you have obtained may not be accurate. Move to an area with
t
emperature between 10ºC to 40ºC (50ºF - 104ºF). Do not artificially heat o
r
cool the meter
B
attery is dead.
C
hange battery now
T
est strip may be damaged o
r
t
est was not performed correctly
P
erform the test with new test strip and follow the test procedure correctly
T
est strip is used
P
erform the test with a new test strip only
N
o responses when strip
i
s inserted into the meter
Maybe:
1
. Battery is dead
2
. Wrong strip is inserted
3. Meter is defective
You have to:
1
.Replace battery
2
.Insert the test strip correctly
3. Contact us for help
No responses when blood
s
ample is applied to the
s
trip
Maybe:
1
.Blood sample is not sufficient
2
.Meter is defective
You have to:
1
.Repeat test with sufficient sample
2
. Perform Meter Quality Check by inserting check strip