DISPLAYING MESSAGES AND TROUBLESHOOTING
As listed below are problems that may appear with your GLM-76 Blood Glucose Meter when
performing a test. These messages will help you to identify certain problems. If you have any
concerns or see any other error display, please contact us or the local distributor for help.
NOTE:
If there is any error message
displayed but not listed here,
or your meter shuts down for
no reason, you can reset the
meter or replace the battery.
isplay check
f some parts of the display do not work Contact your local distributor for
elp
Blood glucose result may be
h
igher than 33.3 mmol/L or 60
g/dL.
If this is not confirmed by the way you feel, review proper testing
procedure and perform a quality check with control solution. Repeat
blood test, if “HI” still appears, call your doctor’s immediately.
Blood glucose result may be
lower than 1.1 /L or 20 mg/dL.
If this is not confirmed by the way you feel, review proper testing procedure
and perform a quality check with control solution. Repeat blood test. If “LO”
still appears, call your doctor’s immediately
emperature is above or belo
he operating range of test strips
The result you have obtained may not be accurate. Move to an area with
t
emperature between 10ºC to 40ºC (50ºF - 104ºF). Do not artificially heat o
attery is dead.
hange battery now
est strip may be damaged o
est was not performed correctly
erform the test with new test strip and follow the test procedure correctly
est strip is used
erform the test with a new test strip only
o responses when strip
s inserted into the meter
. Battery is dead
. Wrong strip is inserted
.Replace battery
.Insert the test strip correctly
No responses when blood
s
ample is applied to the
trip
.Blood sample is not sufficient
.Meter is defective
.Repeat test with sufficient sample
. Perform Meter Quality Check by inserting check strip。
20