8
DCV+ACV +
734 02
Accuracy
Reso-
lution
Range
DC,
10 to
20Hz
DC,
1k to
10kHz
DC,
20Hz to
1kHz
DC,
10k to
20kHz
DC,
20k to
50kHz
DC,
50k to
100kHz
Input
Impe-
dance
Maxi-
mum
Input
Voltage
1000V
rms
AC
1000V
DC
11MΩ
<50pF
10MΩ
<50pF
5V
50V
500V
1000V
0.001V
0.01V
0.1V
1V
1.5+10
*1
*2
0.5+10
*1
2+10
*2
5+20
*2
1+10
*1
*2
---
Maximum Reading 5000, Crest factor: <3
Accuracy
*1: At 5 to 100% of range
*2: At 10 to 100% of range
CMRR: 80dB or more DC to 60Hz (Rs=1kΩ)
Response time: Approx. 5 sec
DC Current Measurement A
Range
500µA
5000µA
50mA
500mA
5A
10A
0.01µA
0.1µA
0.001mA
0.01mA
0.0001A
0.001A
0.2+2
0.6+2
500mA
Protected by a
500mA/600V
fuse.
15A
Protected by a
15A/600V fuse.
<0.11mV/µA
<4mV/mA
<0.1V/A
Resolution Voltage Drop
Maximum
Input Current
Accuracy
73401/02
Response time: 1 sec max.