40
<Table 2>
After completing the calibration for ranges in Table 1, carry out
calibration for “ Frequency Characteristic”.
The calibration for is frequency characteristic required for AC voltage
and AC current ranges (marked with ).
Calibration is carried out at the designated frequency in table 2.
1) Set the calibrator to Input value as an input to the instrument.
2) Press the MEMORY key.
3) After 20 seconds, buzzer sounds and the instrument confirm calibration.
(Do not next key operation till buzzer sounds.)
Table 2. Input Signal for Calibration
AC 500mV
AC 5V
AC 50V
AC 500V
AC 1000V
1kHz(400Hz)
AC 500µA
AC 5000µA
AC 50mA
AC 500mA
AC 5A
AC 10A
500.00
5.0000
50.000
500.00
1000.0
500.00
5000.0
50.000
500.00
5.0000
10.000
mV
V
V
V
V
µA
µA
mA
mA
A
A
Input UnitRange
1kHz
5kHz