30
15. Hygiene and maintenance – General
Follow your local and country-specific laws, directives, standards and guidelines for cleaning, reprocessing and sterilization.
> Wear protective clothing.
> Switch the handpiece off.
> Remove the doriot style disposable prophy angle from the handpiece.
> Pull the nosecone off the handpiece and dispose of the barrier sleeve.
> Clean and reprocess the handpiece and the foot control immediately after every treatment!
> Sterilize the handpiece nosecone after cleaning and reprocessing.
> The handpiece and foot control are not approved for mechanical cleaning (thermo washer disinfector) and sterilization.
Do not submerge the medical device into reprocessing agent or an ultrasonic bath!
Part Number Part Description Guidance
295721 White Handpiece
Follow reprocessing instructions starting on Page 32
For the handpiece nosecone, follow sterilization instructions on Pages 33-34
295734 Foot Controller Follow reprocessing instructions starting on Page 32
295786 Stick Not to be used with contaminated Components. Dispose if contaminated
295788 Charger, including adapter Follow reprocessing instructions starting on Page 32
295789 Handpiece Cradle Follow reprocessing instructions starting on Page 32
295830 Disposable Barrier Sleeve Single Use Only. Dispose after Use