37
Regular checks
Regular servicing of function and safety including the accessories is necessary and should be carried out at least once
every three years, unless shorter intervals are prescribed by law. The inspection must be undertaken by a qualified
organization and must include the following procedures:
Handpiece
> External visual inspection
> Function test
Foot control
> External visual inspection
> Function test with check to see if the maximum speed can be reached
The regular inspection must only be performed by an authorized Young Dental service partner (see page 43).
Defective or disused batteries must only be replaced by an authorized Young Dental service partner (see page 43).
18. Servicing