ZEUS-200S/150 ZO-OM-007
REV : 08 27 ZERONE
7.2 Output
ZEUS-200S
Tolerance : ±20%
Mode output power Carrier Crest Duty Repeat
Freq. Factors Rate Freq.
Pure Cut 200W at 200ohm 400 1.7 100% Continuous
Blend1 150W at 200ohm 400 2.0 67% 33
Blend2 120W at 200ohm 400 2.7 42% 33
Contact coagulation 80W at 200ohm 400 2.6 25% 33
Spray coagulation 80W at 500ohm 400 4.4 17% 33
Bipolar coagulation 80W at 100ohm 500 2.0 100% Continuous
These parameters can be changed without notice.
ZEUS-150
Tolerance : ±20%
Mode output power Carrier Crest Duty Repeat
Freq. Factors Rate Freq.
Pure Cut 150W at 200ohm 400 1.6 100% Continuous
Blend1 120W at 200ohm 400 1.9 67% 33
Blend2 100W at 200ohm 400 2.3 42% 33
Contact coagulation 80W at 200ohm 400 3.0 25% 33
Spray coagulation 60W at 500ohm 400 4.1 17% 33
Bipolar coagulation 80W at 100ohm 500 1.6 100% Continuous
These parameters can be changed without notice.