BLOOD GLUCOSE MONITORING SYSTEM
SYSTÈME DE SURVEILLANCE DE LA GLYCÉMIE
User Guide
For in vitro Diagnostic Use
in Animals
Featuring the AlphaTRAK 2 Meter
Distributed by
Zoetis Inc.
Kalamazoo, MI 49007
www.AlphaTRAKmeter.com
Contact Customer Care at 1-888-963-8471
Question and Answer IconNeed help?

Do you have a question about the Zoetis AlphaTRAK 2 Meter and is the answer not in the manual?

Zoetis AlphaTRAK 2 Meter Specifications

General IconGeneral
BrandZoetis
ModelAlphaTRAK 2 Meter
CategoryBlood Glucose Meter
LanguageEnglish

Summary

AlphaTRAK 2 Meter Overview

Important System Information

Critical Usage Warning: Test Strip Compatibility

Use only AlphaTRAK 2 test strips and control solution for accurate results.

Setting Up the AlphaTRAK 2 Meter

Obtaining a Blood Sample

Performing a Blood Glucose Test

Blood Glucose Test Procedure

Steps to perform a blood glucose test, including site selection and washing.

Test Strip Handling and Information

System Check Screen Verification

System Check Display Confirmation

Verify meter display matches example. Do not use if mismatched. Contact Customer Care.

Coding the Meter

Code Matching Procedure

Use C/m buttons to match code on meter to test strip vial for accurate results.

Blood Application to Test Strip

Blood Sample Acceptance Confirmation

Look for lines or listen for beep indicating enough blood. Add more if needed within 60s.

Viewing Test Results

Result Display and Error Codes

Result appears on display. Higher glucose levels take longer. Note error codes.

Interpreting Blood Glucose Results

Critical Health Indicator Warning

Low/high readings can signal serious medical conditions. Consult veterinarian.

Interpreting Low and High Glucose Results

Low Blood Glucose (Hypoglycaemia) Interpretation

Below 65 mg/dL (3.6 mmol/L). Repeat test, control test, consult vet if symptoms persist.

High Blood Glucose (Hyperglycaemia) Interpretation

Above 250 mg/dL (13.9 mmol/L). Repeat test, control test, consult vet if symptoms persist.

LO Result Display Interpretation

LO Result Meaning, Symptoms, and Action

Result < 20 mg/dL (1.1 mmol/L). Indicates severe low glucose. Symptoms listed for dogs and cats.

HI Result Display Interpretation

HI Result Meaning, Symptoms, and Action

Result > 750 mg/dL (41.7 mmol/L). Indicates severe high glucose. Symptoms listed.

Troubleshooting Error Codes

Error Code Er1: Sample or Meter Issues

Sample too small, strip issue, meter error, or glucose < 20 mg/dL. Follow vet advice/control test.

Error Code Er2: High Glucose or Meter Issues

Strip issue, meter error, or glucose > 750 mg/dL. Follow vet advice/control test.

Error Code Er3: Incorrect Test Procedure

Blood applied before symbols or strip insertion. Wait for symbols, perform control test.

Error Code Er4: Strip or Meter Issues

Problem with test strip or meter. Perform control solution test.

Error Code Er6: Accidental Button Press

m or C button pressed during sample application. Retest without pressing buttons.

Troubleshooting Meter Operation

Meter Fails to Enter Test Mode

Check strip insertion, battery installation, or replace dead battery.

Test Not Starting After Blood Application

Using AlphaTRAK Control Solution

Handling Out of Range Control Solution Results

Causes for Out-of-Range Control Results

Expired solution/strip, incorrect procedure, meter malfunction, wrong code, temperature issues.

Performing a Control Solution Test

Viewing Control Solution Test Results

Accessing Meter Memory

Reviewing Test Results in Memory

Meter Maintenance

Critical Battery Replacement Timing

Replace battery immediately when indicator appears to prevent meter failure.

Cleaning the Meter

Critical Cleaning Restriction: No Immersion

Do not immerse the meter in water or other liquid to prevent damage.

Technical Specifications