Performance Characteristics
34
Within run precision
(100 venous blood tests per glucose level)
Total precision
(100 control solution tests per glucose level)
Concentration
Standard
deviation
(mg/dL)
Coecient of
variation (%)
30~50 mg/dL 2.0 4.2
51~110 mg/dL 3.4 3.6
111~150 mg/dL 4.4 3.2
151~250 mg/dL 6.1 2.5
251~400 mg/dL 6.0 1.6
Concentration
Standard
deviation
(mg/dL)
Coecient of
variation (%)
30~50 mg/dL  1.9 3.9
90~130 mg/dL 4.1 3.8
280~420 mg/dL 9.2 2.6
Regression statistics
Slope 1.0133
Intercept -1.4079 mg/dL
R
2
0.9968
No. of samples 100
Range tested 33.5~445 mg/dL