35
Model PG-101 (2in1. SMART Blood Glucose
Monitoring System)
Test Method Electrochemical sensor
Calibration Method Plasma
Sample Whole blood,
capillary
Sample Size 0.5 μL
Test Time 5 seconds
Memory No limit
Result Range 10~600mg/dL (0.6~33.3mmol/L)
Hematocrit 20-60%
Operating
Temperature
10~40°C (50~104°F)
Operating
Relative Humidity
10-90%
Altitude Up to 3,048 meters (10,000 feets)
Power Source From the smart phone
Size 21.0 mm x 42,7 mm x 8.8 mm
(0.83” x 1.68” x 0.35”)
Weight 4.2 g (0.15 oz)
Specifications
Electrical and safety standards
The meter has been tested for immunity to electro static
discharge as specified in IEC 61000-4-2. The meter has been
tested for immunity to radio frequency interference at the
frequency range and test levels specified in IEC 61000-4-3. The
meter has been tested for electromagnetic emissions specified
in IEC 61326.