Loading...
24
Product
Description
Specifications
(Continued)
Performance
Water
Outlet
Temperature
Arterial
Heat
Exchanger
and
Blanket:
0-42°C.
Cardioplegia:
ice
water
only.
Maximum
Flow
Performance
(
1/8
hp
Pump)
Arterial
Heat
Exchanger*
outlet
port:
5.5
gal/min
(21.0
L/Min)
at
zero
head
13.5
psi
(700
mmHg)
maximum
pressure
Blanket
Supply“
outlet
port:
0.8
gaL’min
(3.0
mm
at
zero
head
11.8
psi
(610
mmHg)
maximum
pressure
Cardioplegia
Water
Supplyeport:
2‘1
gal/min
(7.8
L/Min)
7.5
psi
(336
mmHg)
maximum
pressure
Ice
Supply
If
the
large
tank
is
filled
to
the
maximum
level
with
25°C
water,
the
Ice
Maker
can
provide
a
maximum
of
50
lbs
(22,7
kg)
in
8
hours.
If
the
large
tank
is
filled
to
the
maximum
level
with
3°C
water,
the
Ice
Maker
can
provide
a
minimum
of33.5
lbs
(15,1
kg)
in
3
hours,
Temperature
Probe
Module
operating
range:
0-50°C
Remote
temperature
sensing
and
display
accuracy:
i
03°C
Venous
differential
mode:
gradient
=
rewarm
water
temperature
is
held
to
a
maximum
of
6°C
more
than
the
incoming
venous
blood
temperature.
10°
gradient
=
rewarm
water
temperature
is
held
to
a
maximum
of
10°C
more
than
the
incoming
venous
blood
temperature.
*Note:
Performance
will
be
different
with
both
a
heat
exchanger
and
a
blanket
being
supplied
water
at
the
same
time.
24