If resistance is felt upon removal of the Clip Applier, insert the Dilator (or an 18-gauge thin-wall needle) into the Access Ports, one at a
time, to unlock the Thumb Advancer.
A “CLICK” should be heard to confirm release. Check that the number “3” on theThumb Advancer exits the number window completely.
Pull back the Thumb Advancer.
Slide the Safety Release to collapse theLocator Wings. Check that the number “2” on the Plunger exits the number window completely.
Gently push down on the patient’s body with the left hand. Remove the Clip Applier with the right hand without rocking or twisting.
Use of Access Ports
StarClose SE
Vascular Closure System