Part number 20220, Rev 0
1
PREMIER PHARMACY / VACCINE UNDERCOUNTER FREE STANDING
REFRIGERATORS
PH-ABT-UCFS-0204G, PH-ABT-UCFS-0204, PH-ABT-UCFS-0504G
OWNER’S INSTRUCTIONS
This manual describes how to operate and care for your appliance to get the best, most efficient, performance.
READ THIS BOOK!
Note to Customer:
This merchandise was carefully packed and thoroughly inspected before leaving our plant. Responsibility for its
safe delivery was assumed by the carrier upon acceptance of the shipment. As directed on the side of your
packing carton, claims for loss or damage sustained in transit must be made on the carrier as follows:
A.) Visible Loss, Damage, Shortage External Evidence of Loss or Damage: This type of damage must be noted on
the freight bill and acknowledged by the carrier’s agent (driver) at time of delivery. Make sure you get a signed
copy. Send a written request for an inspection to the carrier.
B.) Concealed Damage: This type of damage may not be discovered until the unit is being unpacked. When
concealed damage is discovered, stop unpacking immediately and contact the carrier immediately to report
the claim and request an inspection. This should be done as soon as possible and, in any case, must be done
within 15 days or receiving the merchandise. If at all possible, do not move the item and save all packaging
material for carrier’s inspection.
C.) FAILURE TO FOLLOW THESE INSTRUCTIONS MAY RESULT IN THE CARRIER REFUSING TO HONOR YOUR
COMPANY’S CLAIM. UNDER NO CIRCUMSTANCES SHOULD THE MERCHANDISE BE RETURNED TO THE
MANUFACTURER. NO RETURNS WILL BE ACCEPTED WITHOUT PRIOR AUTHORIZATION.
Horizon Scientific
125 Varnfield Drive
Summerville, SC 29483
Phone: 800-648-4041
Fax: 843-821-8051
Warranty:
Two-Year Parts and Labor
5 Years Compressor Parts
Revision Date
03/08/2013

Questions and Answers

Summary

Receiving and Shipping Damage Claims

Handling Visible Shipping Damage

Note visible damage on freight bill, get signed copy, send written request for inspection.

Handling Concealed Shipping Damage

Report concealed damage to carrier within 15 days, retain packaging, request inspection.

Installation Instructions

Uncrating and Placement

Move unit close to final location before unpacking. Ensure 4 inches back and 3 inches side air flow.

Leveling the Unit

Use adjustable leveling legs on front corners. Turn counterclockwise to raise, clockwise to lower.

Power Source Requirements

Use dedicated circuit, 20 amp breaker. Plug into 3-prong grounded outlet. Do not use extension cords.

Starting the Refrigerator

Initial Startup Procedure

Plug in unit after leveling. Allow 4 hours for cooling before adding product. Monitor initial cycling.

Shelf Installation Guide

Hook shelf clips onto pilaster strips. Position shelves evenly. Ensure corners are seated properly.

Temperature Control

Factory Preset Temperature Control

Control is factory preset. Contact Technical Support before making adjustments for proper procedure.

Digital Display Functions

Display shows current temp, alarms, highest/lowest recorded temps. Buttons control functions and setpoints.

Normal Mode Operation

Display shows current temp. HI/LO icon for setpoints, Cooling icon for compressor, Defrosting icon for manual defrost.

Program Mode Adjustments

Access program mode via SET button hold. Adjust differential temp, alarm delay, or Celsius/Fahrenheit display.

Refrigerator Operation

Initial Cooling and Loading

Allow 8 hours for initial cooling. Load product in stages to stabilize temperature. Avoid blocking airflow.

Loading Tips for Optimal Performance

Maintain 2-3 inches clearance, avoid blocking fan, don't overload (75% max), distribute load evenly.

Quick Troubleshooting Guide

Unit Does Not Run Troubleshooting

Check power cord, outlet, breaker, GFCI. Ensure correct voltage (110-120V 60Hz).

Temperature Maintenance Issues

Check room temp, door closure, load overload, product placement, and evaporator frost.

Appliance Runs Too Long Troubleshooting

Check for prolonged door openings, control set too cold, or high room temperature.

Warm Exterior Wall Surface

Exterior walls can be up to 30 degrees warmer than room temp due to condenser coils; this is normal.

Compressor Noise Troubleshooting

Check unit level, pipe contact, compressor overheating (room temp 65-85F), or voltage supply.

Moisture Collection Inside Unit

Check door gasket seal, items blocking door, or excessive door openings. Ensure drain tube water trap.

Moisture Collection on Outside Surface

Hot, humid weather increases condensation. As humidity decreases, moisture will disappear.

Display Errors and Compressor Status

Troubleshoot display codes like 'SO', 'SC', 'HH', 'LL' by checking temperature sensors or ambient temp.

Maintenance and Cleaning

Cleaning Interior and Door Liners

Use soap/water or baking soda solution. Wring out cloth before cleaning electrical parts.

Cleaning Glass Doors

Clean glass with mild detergent and water on a soft cloth or sponge. Rinse and wipe dry.

Cleaning Door Gaskets

Wipe door gaskets and their seating surfaces with a clean soft cloth.

Cleaning Shelves

Use soap and water. Do not wash removable shelves in the dishwasher.

Cleaning Exterior and Handles

Use soap/water or non-abrasive glass cleaner. Avoid commercial cleaners, ammonia, or alcohol.

Product Warranty Information

Horizon Scientific, Inc. Product Warranty

Two-year warranty covers unit, cabinet, and parts against defects when installed and used correctly.

Additional Three Year Compressor Warranty

Covers hermetically and semi-hermetically sealed compressor parts for an additional three years (five total).

Warranty Limitations and Exclusions

Excludes damage from alteration, neglect, abuse, misuse, accident, transit, fire, flood, or external causes.

Warranty Claims Procedure

Contact manufacturer at 800-648-4041. Provide model, serial number, proof of purchase, and defect details.